Advertisement

NurseCall N46 Wall Transmitter
5.7

General configuration

The * marks a factory setting.
Bridge State
B3
B4
B5
B6
B7
B8
(o)
Presence Processing activated (B6 open). Models that do not feature the output relays are
delivered with the Staff Presence Processing disabled (B6 soldered).
Bosch Security Systems
Function
Open*
Technical Alarms are not repeated.
Soldered
Technical Alarms are repeated until alarm acknowledgement
(max 20 min.)
Open*
Calls (except technical alarms) are repeated until call acknowledge-
ment (max 20 minutes)
Soldered
Calls (except technical alarms) are not repeated.
Open*
A daily message is sent every 24 hours.
Soldered
Daily messages are suppressed.
(o)
Staff Presence Processing is activated.
Open
(o)
Staff Presence Processing is disabled.
Soldered
Open*
Activation of J6 [4-5] terminal input causes a Call for Help 2.
Soldered
Activation of J6 [4-5] terminal input causes a Call for Assistance.
Open*
The Assistance Call does not depend on the presence of the Care Per-
sonnel.
Soldered
The Assistance Call is possible only if the Care Personnel has signaled
his/her presence.
NurseCall N46 Wall Transmitters that feature the output relays are delivered with the Staff
User Manual
Installation | en
19
970.030 | v1.0 | 2009.06