Call Acknowledgement (Staff Presence); Call For Help 2 (Reserve); Technical Alarm; Caregiver Arrival - Bosch NurseCall N46 User Manual

Wall transmitter
Hide thumbs Also See for NurseCall N46:
Table of Contents

Advertisement

NurseCall N46 Wall Transmitter
6.3

Call Acknowledgement (Staff Presence)

The call acknowledgement is triggered by pushing the Acknowledgement button (green).
The call acknowledgement stops immediately all repetitions of the calls.
Optionally, a call acknowledgement triggers the Staff Presence signal:
6.4

Call for Help 2 (reserve)

The Call for Help 2 is triggered by pushing the external button connected to terminals 4 and 5
of the terminal block J6.
The call can be triggered by closing (NO) or opening (NC) the contact.
The Call for Help 2 is repeated about every 2 minutes until a call acknowledgement takes
place, up to a maximum of 20 minutes. This repeat function can be disabled by setting the sol-
der bridge B4.
The Call for Help 2 is signaled in the same way as the Call for Help on the signal lamp and the
relays.
6.5

Technical Alarm

The Technical Alarm is triggered by activating the contact connected to terminals 3 and 4 of
the terminal block J6.
The alarm can be triggered by closing (NO) or opening (NC) the contact 3.
Normally, the Technical Alarm is not repeated, but it is possible to activate the repetition of
the technical alarms by setting the solder bridge B3. In this case, the Technical Alarm is
repeated about every 2 minutes until a call acknowledgement takes place, up to a maximum of
20 minutes.
The Technical Alarm is signaled in the same way as the Call for Help on the signal lamp and the
relays.
6.6

Caregiver Arrival

The Caregiver Arrival message is triggered when the Caregiver inserts the coded key in the
connector on the side of the housing (J7), when entering the room. This message will appear
if the optional coded key function is available on the device.
Up to four categories of Caregivers may be identified by using different keys.
The Caregiver Arrival message is not repeated.
Bosch Security Systems
by green flashes on the signal lamp and by activating the Staff Presence relay when
the Caregiver pushes the Acknowledgement button when entering the room.
by switching off the lamp and releasing the Staff Presence relay when the Caregiver
pushes the Acknowledgement button and leaves the room.