Chapter 7 Tutorial
Relay Life and Preventive Maintenance

Relay Life

As a relay is used, the contacts begin to wear and the resistance of the
closed contacts increases. The initial contact resistance of a relay is
typically 50 m (plus lead resistance). When the contact resistance
exceeds 20 to 50 times its initial value, the contact resistance becomes
very erratic and the relay should probably be replaced. For most
applications, a relay with contact resistance greater than 1 should be
replaced. The graph below shows the typical contact resistance
characteristics of the relays used on the 34970A/34972A switching
modules.
Typical Relay Lifetime
Full Load
1
0
1k
10k
100k
Relay Operations (Cycles)
Relay Load
For most applications, the load switched by the relay is the most
important factor affecting relay life. As shown in the graph below, relay
life is maximized by switching low power. As the power switched
increases, relay life deteriorates.
Maximum Switching Capacity
300V
240V
180V
120V
60V
0.2A
0.4A
0.6A
10 mA 20 mA 30 mA 40 mA 50 mA
Current Switched
322
Typical Signals No Load
1M
10M
100M
Armature (34901A, 34903A, 34904A, 34908A)
0.8A
1A
Reed (34902A)