Agilent Technologies 34972A User Manual page 125

Data acquisition / switch unit
Table of Contents

Advertisement

• The thermocouple check feature allows you to verify that your
thermocouples are properly connected to the screw terminals for
measurements. If you enable this feature, the instrument measures
the channel resistance after each thermocouple measurement to
ensure a proper connection. If an open connection is detected (greater
than 5 k on the 10 k range), the instrument reports an overload
condition for that channel (or displays "OPEN T/C" on the front
panel).
• Front-Panel Operation: To select the thermocouple function on the
active channel, choose the following items.
TEMPERATURE , THERMOCOUPLE
To select the thermocouple type for the active channel, choose the
following item.
J TYPE T/C
To enable the thermocouple check feature on the active channel
(opens are reported as "OPEN T/C"), choose the following item.
T/C CHECK ON
To select the reference junction source for the active channel, choose
one of the following items.
INTERNAL REF , EXTERNAL REF , FIXED REF
Note: Before selecting an external source, be sure to configure
channel 01 in the lowest slot for a thermistor or RTD measurement.
Chapter 4 Features and Functions