Chapter 2 Front-Panel Overview
To Configure the Remote Interface - 34970A
RS-232 Configuration
1 Select the RS-232 interface.
RS-232
2 Select the baud rate.
Select one of the following: 1200, 2400, 4800, 9600, 19200, 38400,
57600 (factory setting), or 115200 baud.
19200 BAUD
3 Select the parity and number of data bits.
Select one of the following: None (8 data bits, factory setting), Even
(7 data bits), or Odd (7 data bits). When you set the parity, you are also
indirectly setting the number of data bits.
EVEN, 7 BITS
4 Select the flow control method.
Select one of the following: None (no flow control), RTS/CTS, DTR/DSR,
XON/XOFF (factory setting), or Modem.
FLOW DTR/DSR
5 Save the changes and exit the menu.
54