desired ramp rate (PR). Dwell segments (PD) can be
attached to each program level (PL) to hold that tempera-
ture for a specified number of minutes.
Multi-Programmable Controller Program Entry
(Models F46120CM, F46120CM-33, F46120CM-33-75,
F46120CM-75, F46128CM, F46128CM-75, F46240CM,
F46240CM-33, F46240CM-33-75, F46240CM-75,
F46248CM, F46248CM-75)
The multi-programmable controller in the F46120CM-75,
F46128CM-75, F46240CM-75 and F46248CM-75 units
provides up to 15 separate programs of 8 ramps and 8
dwells each. The F46120CM, 46128CM, F46240CM and
F46248CM models have four separate programs These
controllers allow you to link programs together. These
functions are governed by the controller's first two program-
ming parameters, "Pnr" and Cnt."
To select program number
Push scroll button until "Pnr 1" is displayed. Push the up or
down button to select a program number from 1 to 15. On
the F46120CM, F46120CM-33, F46128CM, F46240CM,
F46240CM-33 and F46248CM models, this is selected
between 1-4.
To link programs together
Push scroll button until "Cnt n" is displayed. Press and
release the up and down buttons to switch between
"Cnt y" (continue yes) and "Cnt n" (continue no). The effect
of selecting "Cnt y" is to continue the program to the next
program number. For example, if in program #3 you select
"Cnt y," when program #3 is complete, program #4 will run
automatically. Setting "Cnt y" in program #15 will initiate the
start of program #1 upon the completion of program
#15.Each program will complete the selected number of
loops before continuing (see Loop Count).
These two parameters ("Pnr" and "Cnt") are the only ones
that differ from the single program 8 ramp and 8 dwell
control. Program entry for the multi-programmable models
continues with the following instructions, which are appli-
cable to both the multi-programmable and single program-
mable models.