Blink
FAULT
code
4
Power time
exceeded
5
Brake failure
6
No neutral position
when switching
Scooter on.
7
Fault in speed
potentiometer
Consequence
for the Scooter
Stops driving
The Scooter has used up too much
power for a long period, probably
because the motor has been overloaded
or the Scooter has attempted to work
against insurmountable resistance.
Switch off the Scooter, wait a few
minutes and then switch it on again.
The electronic system has determined a
motor short-circuit. Check the wiring
harness for short-circuits and check the
motor: Contact your Invacare® dealer.
Stops driving
Ensure that the disengaging lever is in
the engaged position.
There is a defect in the braking coil or in
the cabling. Check the wheel lock and
cabling for open or short-circuited
circuits. Contact your Invacare® dealer.
Stops driving
Drive lever is not in neutral when the
keyswitch was turned. Put the drive lever
in neutral, turn the power off and then
turn on again.
It may be necessary to recalibrate the
drive lever. Contact your Invacare®
dealer.
Stops driving
The drive lever electronics could be
faulty or incorrectly connected. Contact
your Invacare® dealer.