To remove the Power Cartridge, push up on the release lever on the Power Cartridge. The
Power Cartridge will slide out as you grasp it for removal.
power cartridge duration-of-Use: The duration-of-use of the Power Cartridge is displayed
at the top of the user control panel. A variety of factors, such as flow setting, Pulse or Con-
tinuous Flow Mode will impact the duration-of-use. The following table provides duration-of-
use estimates for the Eclipse operating on a fully charged Power Cartridge given certain flow
settings and operating conditions.
typIcal nEW poWER caRtRIdgE
dURatIon-of-USE
floW SEttIng
0.5
1.0
2.0
3.0
4.0
5.0
6.0
notE: Times will degrade with Power Cartridge use, over time, and the increasing age
of the Power Cartridge.
caUtIon
poWER caRtRIdgE
contInUoUS floW
modE (lpm)
4.4 hours
3.7 hours
2.4 hours
1.3 hours
-
-
-
StoRE thE poWER caRtRIdgE In a cool and
dRy locatIon. thIS WIll hElp to aSSURE thE
longEvIty of yoUR poWER caRtRIdgE.
pUlSE floW modE
(SEttIng)
-
4.4 hours
3.6 hours
3.0 hours
2.6 hours
2.3 hours
2.1 hours