Caution
Be sure ambient temperature does not
exceed 40°C (104°F). The recommended
ambient temperature is 17°C - 27°C.
Ambients above this level may result in
damage to the controller.
Caution
Allow at least six inches of space
between the furnace and any combustible
surface. This permits the heat from the
furnace case to escape so as not to cre-
ate a possible fire hazard.
Warning
To avoid electrical shock, this furnace
must always use a properly grounded
outlet of correct voltage and current han-
dling capacity.

Installation

Site Selection

Install furnace on a sturdy surface and allow
adequate space for ventilation.

Electrical Connections

The electrical ratings are located on the spec-
ification plate on the back of the furnace.
Consult Barnstead International if your electri-
cal service is different than those listed on the
specification plate. Be sure the front power
switch is in the OFF position before connect-
ing the furnace to your electrical supply.