Barnstead 47900 Operation Manual page 22

Series 1165, 1166 & 1205
Table of Contents

Advertisement

8 S
& 4
16 S
EGMENT
X
22
P
EGMENT
ROGRAMMABLE
M
/OTP
ODELS W
PrG: Program number (displayed when a pro-
gram is running; 4x16 programmable models
only.)
IdHi: Deviation High Alarm
tunE: One-shot autotune enable.
run LiSt (Program Run List)
PrG: Currently running program (only used on
4x16 programmable models)
StAt: Displays the program status [OFF, run
(running active program), hoLd (program on
hold), HbAc (waiting for process to catch up),
End (program completed)] in the lower dis-
play. The controller will default to "OFF."
FASt: Fast run through program (no/YES).
The controller will default to "no."
SEG.d: Flash active segment type in the
lower display of the home display (no/YES).
The controller will default to "no."
ProG LiSt (Program Edit List)
PrG.n: Press the UP or DOWN ARROW to
select the program number (program number
will be displayed in lower display on 4x16 pro-
grammable models only.)
Hb: Press the UP or DOWN ARROW to
select the holdback type [OFF (disables hold-
back), Lo (deviation low holdback), Hi (devia-
tion high holdback) or bAnd (deviation band
holdback)] for the entire program. The con-
troller will default to "OFF."
Hb.U: Press the UP or DOWN ARROW to
select the holdback value (in display units).

Hide quick links: