Fluke 1750 Calibration Manual page 17

Power recorder
Hide thumbs Also See for 1750:
Table of Contents

Advertisement

7. Connect the calibrator NORMAL HI to inputs A, B, C, and J of the Recorder.
8. Connect the NORMAL LO to input N and the protective conductor terminal . of
the Recorder.
9. Apply the voltages from Table 4 and list the Recorder readings for channels AN, BN,
CN, and NG in the appropriate areas. Sine waves will be displayed for AN, BN and
CN, all at the same amplitude and phase shift. NG will be inverted.
If the voltage goes over scale, right click on the voltage scale on the left
margin of the graphical display, then click on the Zoom to Fit box to
resize.
Applied
60 Vrms @ Local Hz
240 Vrms @ Local Hz
600 Vrms @ Local Hz
10. Verify that A, B, C, and J LEDs are on. N should be off.
11. Set the calibrator to STBY.
12. Verify that A, B, C, and J LEDs are off.
Figure 1. Power View Software Live View
Note
Table 4. AC Voltage Accuracy Verification
AN Reading
BN Reading
power analyze live view screen.bmp
CN Reading
NG Reading
Power Recorder
Verification Tests
Limits
± 1.12 V
± 1.48 V
± 2.20 V
9