Fluke 1750 Calibration Manual page 15

Power recorder
Hide thumbs Also See for 1750:
Table of Contents

Advertisement

The cables listed in Table 3 are necessary for verification and calibration adjustment. To
assemble the cables, see cable assembly figures in the Test Cables section.
Cable
Ethernet cable
Voltage calibration cable
with BNC-banana adapter
Four jumper cables, safety-
banana type
Current calibration cable
Rogowski resistor
measurement cable
Rogowski calibration cable
Ground calibration cable
Fluke recommends the use of Pomona brand cables and banana plugs when building these cables.
Table 3. Required Cables
Required Characteristics
CAT 5 with RJ-45 connectors
Double banana to safety banana
and spade lug
Shielded cable with double banana
to CT pins in Redel connector with
100 A CT ID resistor
Safety banana to Redel
Shielded cable with double banana
to Rogowski pins in Redel connector
with 100 A Rogowski ID resistor
Lug to banana
Required Equipment
Notes
Supplied with the Recorder
See Figure 5
See Figure 6
See Figure 4
See Figure 3
See Figure 2
Power Recorder
7

Advertisement