U1270 Series Service Guide
Table 1-3 Adjustment input values (continued)
Test function
Step
30 mA
300 mA
ACmA/ACA
3 A
10 A
OPEN
10 nF
100 nF
1000 nF
Capacitance
10 μF
100 μF
1000 μF
10 mF
OPEN
Calibration Procedures
Using the Front Panel for Adjustments
Reference value
Valid reference input
0.9 to 1.1 × Reference value
03.000 mA (70 Hz)
0.9 to 1.1 × Reference value
30.000 mA (70 Hz)
0.9 to 1.1 × Reference value
030.00 mA (70 Hz)
0.9 to 1.1 × Reference value
300.00 mA (70 Hz)
0.9 to 1.1 × Reference value
0.3000 A (70 Hz)
0.9 to 1.1 × Reference value
3.0000 A (70 Hz)
0.9 to 1.1 × Reference value
3.0000 A (70 Hz)
0.9 to 1.1 × Reference value
10.000 A (70 Hz)
With cables connected to
OPEN
CAP/COM terminals
0.9 to 1.1 × Reference value
04.000 nF
0.9 to 1.1 × Reference value
10.000 nF
0.9 to 1.1 × Reference value
010.00 nF
0.9 to 1.1 × Reference value
100.00 nF
0.9 to 1.1 × Reference value
0100.0 nF
0.9 to 1.1 × Reference value
1000.0 nF
10.000 μF
0.9 to 1.1 × Reference value
100.00 μF
0.9 to 1.1 × Reference value
1000.0 μF
0.9 to 1.1 × Reference value
0.9 to 1.1 × Reference value
10.000 mF
OPEN terminals (no cables
OPEN
connected to CAP/COM
terminals)
1
[1]
45