8.10.4 Setting Impulse Response
The Impulse Response option enables you to force the meter to transmit in
one direction only, without changing the AGC setting. You can then
diagnose problems with transducer, pipe, or fluid configurations. To enter
the option, scroll down to the Impulse Response entry in the Service Menu
and press
[ENTER]
Stopped.
.
Transmit
The first prompt asks in which direction you wish the meter to transmit,
1.
upstream or downstream.
Press
[ENTER]
a.
Use the [
b.
Press
[ENTER]
c.
Transport® Model PT878 Portable Liquid Flowmeter User's Manual
. The screen appears similar to Figure 127 below.
Impulse Response
Direction
Percent Gain
Figure 127: Impulse Response Option
to open the drop-down list.
] or [
] arrow keys to scroll to the desired number.
to confirm your selection.
Chapter 8. Servicing the PT878
Upstream
Stop
20
%
Exit