Accuracy adders for use in RF Fields: In an RF field of
3 V/m, change the accuracy specifications as follows:
For AC Volts Measurement, add 0.25 % of range
For DC Current Measurement, 30.000 mA range,
add 0.14 % or range
For DC Current Output, add 0.32 % of span
Accuracy for all meter functions is not specified in RF
fields > 3 V/m.
Relative humidity: 95 % up to 30 C, 75 % up to 40 C,
45 % up to 50 C, and 35 % up to 55 C
Vibration: Random 2g, 5 to 500 Hz
Shock: 1 meter drop test
Safety: Complies with EN61010, ANSI/ISA S82.01-1994
and CAN/CSA C22.2 No. 1010.1-92 Overvoltage
Category III.
,
,
Certifications:
Power requirements: Four AA batteries (alkaline
recommended)
Size: 10.0 cm X 20.3 cm X 5.0 cm (3.94 in X 8.00 in X
1.97 in)
Weight: 610 g (1.6 lbs)
ProcessMeter