Manually Stepping the mA Output

To manually step current output:
H
I
Use
or
to step the current up or down in
25 % steps.
J
Touch either
to go to 0 %, or
100 %.

Auto Ramping the Output

Auto ramping can continuously apply a varying stimulus
from the Calibrator to a transmitter, while your hands
remain free to test the response of the transmitter.
L is pressed
When
, the Calibrator produces a
repeating 0 % - 100 % - 0 % ramp in a choice of three
ramp waveforms:
E
0 % - 100 % - 0 % 40-second smooth ramp
P
0 % - 100 % - 0 % 15-second smooth ramp
N
0 % - 100 % - 0 % Stair-step ramp in 25 %
steps, pausing 5 seconds at each step. Steps are
listed in Table 7.
42
G
to go to
To exit ramping, press any button.
Table 7. mA Step Values
Step
0 %
25 %
50 %
75 %
100 %

Storing and Recalling Setups

Up to eight settings can be stored in a nonvolatile
memory to recall for later use. A low battery condition or
a battery change does not jeopardize the stored settings.

Store a Setup

To store a setup:
1.
Create the desired setup.
Push S. The right side of the display changes to
2.
show SAVE SETUP and SAVE DATA.
Push Sto select SAVE SETUP.
3.
Push Y or Z to select the desired memory location (at
4.
the top of the LCD).
Push S to enter the setup.
5.
4 to 20 mA
4.000
8.000
12.000
16.000
20.000