Shut Down Mode
The Calibrator comes with a shut down mode set for 30
minutes (displayed for about 1 second when the
Calibrator is initially turned on). When shut down mode is
enabled, the Calibrator automatically shuts down after
the elapsed time from when the last key was pressed.
Press C until SHUT DOWN appears on the
1.
display.
Use X and W to increase or decrease the time.
2.
Press S to save the setting.
3.
CJC
Cold Junction Compensation (CJC) is a value for the cold
end of a thermocouple at the Meter's end.
Press C until SELECT CJC appears on the
1.
display.
Y
Z
2.
Use
and
to select ON or OFF.
Press S to save the setting.
3.
Celcius and Fahrenheit ( C and F)
Press C until SELECT UNIT C (or F) appears
1.
on the display.
Use X and W to select C or F.
2.
Press S to save the setting.
3.
Multifunction Process Calibrator
Frequency Pulse Output Voltage
Press C until FREQ OUTPUT V Adjust appears
1.
on the display.
Y
Z
Use X, W,
2.
and
to adjust the frequency pulse
output voltage from 1 to 20 V.
Press S to save the setting.
3.
Pulse Output Frequency
Press C until PULSE OUTPUT Hz FREQ Adjust
1.
appears on the display.
Y
Z
Use X, W,
2.
and
to adjust the pulse output
frequency from 2 CPM to 15 kHz.
Press S to save the setting.
3.