Cleaning; Testing The Fuses; Replacing The Battery And Fuses - Fluke 175 User Manual

True rms multimeters
Table of Contents

Advertisement

Models 175, 177 & 179
Users Manual

Cleaning

Wipe the case with a damp cloth and mild detergent. Do not use
abrasives or solvents. Dirt or moisture in the terminals can affect
readings.

Testing the Fuses

X W
To avoid electrical shock or injury, remove the test
leads and any input signals before replacing the fuse.
Test fuses as shown below.
440 mA
<12
HOLD
MIN MAX
RANGE
V
10
Warning
11 A
OK
OK
<.5
OK
OK
AIK12F.EPS

Replacing the Battery and Fuses

XW
To avoid shock, injury, or damage to the Meter:
Use ONLY fuses with the amperage, interrupt,
voltage, and speed ratings specified.
Replace the battery as soon as the low battery
b
indicator (
) appears.
F1
F2
F1 Fuse, 440 mA, 1000 V, FAST
F2 Fuse, 11 A, 1000 V, FAST
B1 Battery, 9 V Alkaline,
NEDA 1604 / 1604A
Warning
b
B1
AIK13F.EPS
Fluke PN 943121
Fluke PN 803293
Fluke PN 614487