Probe Auto Zero

Description
Command Syntax
Example
ADP30X-OM-E Rev B
The PRx:AUTOZERO command initiates an auto zero cycle in the
ADP30X probe to remove any offset drift.
The probe input can remain connected to the test circuit during the auto
zero cycle.
<channel>:AutoZero
<channel>:= {PR1, PR2, PR3, PR4}
The following command initiates an auto zero in the ADP30X
Differential Probe attached to channel 1:
CMD$="PR1:AZ":CALL IBWRT(SCOPE%,CMD$)
ADP30X Operator's Manual
PRx:AUTOZERO, PRx:AZ