Indication Of Alarm Status - Haier DW-86L288 Service Manual

Ultra-low temperature refrigerator
Hide thumbs Also See for DW-86L288:
Table of Contents

Advertisement

"High  temperature"  indicating  lamp:  if  the  lamp  lights,  indicating  the  alarm  display  when  the 
temperature inside the refrigerator is higher than the setting value. 
"Low  temperature"  indicating  lamp:  if  the  lamp  lights,  indicating  the  alarm  display  when  the 
temperature inside the refrigerator is lower than the setting value. 
"Over  voltage"  lamp  lights,  indicating  the  voltage  is  lower  than  184V  (95V)  or  higher  than  253V 
(136V), buzzing. 
"Too  high  ambient  temperature"  indicating  lamp:  if  the  lamp  lights,  indicating  the  ambient 
temperature is higher than 32 . 
"Sensor fault" indicating lamp: if the lamp lights, indicating the sensor has faulted. 
"Power interruption" indicating lamp: if the lamp lights, indicating the interruption of power line. 
"Dirty condenser" indicating lamp: if the lamp lights, indicating the condenser is blocked by dirt, and 
needs to be cleaned. When it is felt that the difference between the condenser and the 
ambient  temperature  is  greater  than  or  equals  13   (after  5  minutes  later),  "dirty 
condenser" indicating lamp lights, accompanied by the buzzer alarm. 
"Low  battery"  indicating  lamp:  if  the  lamp  lights,  indicating  the  low  battery  and  will  be  charged; 
when the battery is connected to AC power, the low temperature refrigerator will charge 
it automatically. 
 
well as the high and low temperature alarm information. 15 minutes after light flashing, 
the buzzer begins to send alarm by voice; for the alarm information on power interruption 
and  overvoltage,  the  buzzer  beings  to  send  alarm  by  voice  after  the  light  flashes  for  1 
minute.  Under  the  power  interruption,  the  display  appears  black  screen,  without  any 
other  information  displayed, and  the  alarm  lamp  operates  under  alarm  mode.  After  the 
black screen lasts for 30 seconds, the display panel will normally display for 5 seconds, and 
alternatively display as per this mode. For alarm information on dirty condenser, when the 
light flashes, the buzzer and the voice will send alarm at the same time; for that on low 
battery,  light  flashing  alarm  is  available,  without  buzzer  alarm;  for  that  on  too  high 
 

Indication of alarm status

  When all alarms appear, light alarm signal will immediately flash to alarm as 
 