Faq; Common Faults Analysis And Maintenance Measures - Haier DW-86L288 Service Manual

Ultra-low temperature refrigerator
Hide thumbs Also See for DW-86L288:
Table of Contents

Advertisement

 
Question: Displayed temperature is not the same as actual test temperature:   
Answer: As the displayed temperature is the temperature at one point in refrigerator, while a 
user  is  testing  the  temperature  of  another  point  in  refrigerator,  as  position  is  different, 
temperature  must  be  different,  meanwhile,  the  user's  testing  tool  may  have  certain 
measurement error with our temperature detector.   
 
Question: The temperature difference between upper part and lower part in refrigerator is 
too large: 
Answer: As our machines start refrigerating from up to down, the upper part temperature is 
lower  than  lower  part,  meanwhile,  the  temperature  difference  is  large.  The  temperature 
difference  of  like  products  in  the  world  is  5~8 
maintained around the temperature. In addition, if a user open and close refrigerator doors 
frequently, big temperature difference between upper part and lower part may appear. 
 
Question: What is voltage‐increase start? 
Answer:  To  ensure  the  normal  operation  of  machine,  our  machines  have  the  functions  of 
increasing and decreasing voltage. When a user's voltage is lower than 183 V, the machine is 
unable to start, at that time, the user needs to check if his line is normal or not; 
 
Question: A user has set ‐86 , why it rises up to ‐82
Answer:  To  ensure the service life of machine, our products are designed as the followings: 
when ambient temperature is higher than 32 , if setting temperature is lower than ‐82 , it 
will return to ‐82 , if ambient temperature is lower than 32 (including 32 ), it will return 
to original setting; 
 
 
 
 