Perfusion refrigerant 
high  temperature  level  perfusion,  the  machine  should  be  in  a  stopped  state 
before  perfusion  R134A  refrigerant  and  then  perfusion  R404A  refrigerant. 
Perfusion was performed in accordance with the attached table! 
Note  
1) To ensure the refrigerant once reperfusion in place, the above refrigerant is 
required when the cylinder is inverted perfusion liquid perfusion in Figure 5 
2) Low‐level refrigerant reperfusion, the same machine should be in a stopped 
state, first perfusion HRA refrigerant, then perfusion HRB refrigerant, because 
of higher the HRB refrigerant pressure. Perfusion was performed in accordance 
with the attached table! 
Irrigation  after  the  infusion  process  tube  sealing,  sealing  effect  is  shown  in 
Figure 6 
Note  
1)  Low  Temperature  perfusion  must  be  accurate,  small  differences  will  affect 
the cooling effect 
2)  low  temperature  and  high  pressure  refrigerant,  refrigerant  cylinder  is 
relatively small, perfusion, to open the cylinder valve slowly to avoid the waste 
of refrigerant. 
3) HRB refrigerant pressure is too high, the general cylinders can not withstand 
the high pressure, so the use of oxygen bottles filled irrigation HRB refrigerant 
cylinder  below  the  export  blocked  with  screws  to  prevent  the  leakage  of 
 
 
  56 / 88 