Haier DW-86L288 Service Manual page 64

Ultra-low temperature refrigerator
Hide thumbs Also See for DW-86L288:
Table of Contents

Advertisement

 
1.E0 alarm 
Ring temperature sensor input voltage 
≥  4.9V,  the  sensor  open;  ≤  0.1V,  the 
short‐circuit 
2.E1 alarm 
When  the  condenser  sensor  input 
voltage  ≥  4.9V,  the  sensor  open;  ≤ 
0.1V, the short‐circuit 
3.E2 alarm 
Main  sensor  input  voltage  ≥  4.9V,  the 
sensor open; ≤ 0.1V, the short‐circuit 
4.E3 alarm 
Heat exchanger sensor input voltage ≥ 
4.9V,  the  sensor  open;  ≤  0.1V,  the 
short‐circuit 
5.  The  condenser 
Condenser 
dirty, alarm 
condenser  temperature  minus  the 
difference  in  ambient  temperature  ≥ 
13  °  C  (for  five  minutes),  alarm 
occurred 
6.  Battery  low 
≤  10.5V  battery  power,  battery  low 
alarm 
alarm 
 
 
probe 
perception 
of 
 
  64 / 88 
 
1.Check  the  ambient  temperature  sensor  terminal  plug  bad  or 
falling; sensor for damage 
2 Check the display board sensor connector terminals are loose or 
poor contact; whether the display panel damage 
1.Check  the  ambient  temperature  sensor  terminal  plug  bad  or 
falling; sensor for damage 
2 Check the display board sensor connector terminals are loose or 
poor contact; whether the display panel damage 
1.Check  the  main  sensor  terminal  plug  bad  or  falling;  sensor  for 
damage 
2  Check  the  main  control  board  sensor  plug  terminal  is  loose  or 
poor contact; master plate for damage 
1.Check  the  heat  exchanger  sensor  terminal  plug  bad  or  falling; 
sensor for damage 
2  Check  the  main  control  board  sensor  plug  terminal  is  loose  or 
poor contact; whether the display panel damage 
3. Replacement spare sensor 
4.  "E3",  long  press  to  "beep  cancel"  after  5  seconds,  "E3"  alarm 
will disappear when the alarm. The temperature stage one minute 
after the start of the Low Temperature startup 
1.Open the front grille and clean the filter 
2.Check  whether  the  condenser  probe  away  from  the  condenser 
outlet too close 
3.The condenser probe take Changuo processing tape 
1.Investigation and whether the battery have expired 
2  the  troubleshoot  battery  switch  on  terminal  block  whether  the 
patch well or switch is damaged 
3.Investigation  and  charging  circuit  is  normal:  strong  power  after 
five  minutes  on  the  refrigerator,  the  test  master  board  battery 
terminal voltage output, the output voltage should be ≤ 5V 