4
Specifications
Supplemental measurement specifications
Table 4-7 Supplemental measurement specifications
DC Voltage
• Measuring Method:
•
Sigma Delta A-to-D converter
• Input Resistance:
•
>10 GΩ ± 2% range (Selectable 100 mV, 1 V ranges)
•
10 MΩ ± 2% range (typical)
• Input Protection:
•
1000 V on all ranges (HI terminal)
Resistance
• Measurement Method:
•
2-wire or 4-wire Ohms
• Open-Circuit Voltage
•
Limited to < 2.8 V
• Input Protection:
•
1000 V on all ranges (HI terminal)
DC Current
• Shunt Resistance:
1 Ω for 10 mA, 100 mA
•
90 Ω for 100 μA, 1 mA
•
0.01 Ω for 1 A, 10 A
•
• Input Protection:
•
Externally accessible at Rear Panel 0.4 A, 500 V fuse for I terminal
•
Internal 11 A, 1000 V fuse for 10 A terminal
108