10. BEFORE ASKING FOR REPAIR

Before asking for repair, refer to the table below.
When
Nothing appears on the
display, even when the scale is
turned on.
[----] remains on the display
and does not change.
The difference between an
estimated weight and the
actual weight is large.
The switches do not function.
The display does not change
even when the switches are
pressed.
[E-1] , [E-2] or [E-4] appears on
the display.
The buzzer remains sounding.
[ E ] appears.
[ -E ] appears
English
Check whether the AC adapter or batteries are installed
correctly.
Check whether the batteries are drained.
Check that nothing is placed on the scale.
Check that nothing is touching the scale.
Check that the scale is leveled.
Check that nothing is touching the scale.
Disconnect the AC adapter and connect it again. Then,
turn the power on.
Remove the batteries and insert them again. Then, turn
the power on.
Ask for repair.
Check whether the ON/OFF switch is left pressed.
This error appears when the object on the scale is
beyond the weighing capacity. Remove the object from
the scale.
This error appears when the weight value is less than 3
kg.
Check that nothing is placed on the scale and nothing is
touching the scale. Then, turn the scale off and turn it
on again.
- 17 -
What to do
AD-6121A