4~20Ma Loop Current As % Readout - UNI-T UT71C Operating Manual

Hide thumbs Also See for UT71C:
Table of Contents

Advertisement

Note
l
Place the Meter in an environment of 18ºC~23ºC
otherwise false reading may be obtained especially
in testing low temperature.
l
The included point contact temperature probe can
only be used with temperature 230ºC below.
l
When temperature measurement has been
completed, remove the temperature probe away from
the multi-purpose socket, and remove the multi-
purpose socket away from the Meter.
I. 4~20 mA loop current as % readout
It shows the mA measured value or output level in %,
in a 4-20mA scale
To use 4~20mA Loop feature, connect the Meter as
follows:
1.
Set the rotary switch to
mA
button to select (4~20mA)% feature.
44
2.
3.
, and press BLUE
Model UT71C/D/E: OPERATING MANUAL
The rest procedure, please follow B. Measuring
Current: DC current measurement (Figure 3-2).
When the readings obtained is:
l
< 4mA, the primary display shows LO
l
l4mA, the primary display shows 0%. ....
20mA, the primary display shows 100%
l
> 20mA, the primary display shows HI