6063B
"CAL 1"
"CAL:INIT 240,60"
"CAL:SAVE"
"DIAG:CAL 0,6063"
"DIAG:CAL 1,16901"
"DIAG:CAL 2l,0"
"DIAG:CAL 22,0"
"DIAG:CAL 23,1"
"DIAG:CAL 26,1"
"DIAG:CAL 27,1"
"DIAG:CAL 28,5000"
"DIAG:CAL 29,5200"
"*RST"
"CURR:SLEW 0.167E6"
"*SAV 0"
"CAL 0"

Disassembly Procedures

The following disassembly procedures are listed in alphabetical order. Before proceeding with any disassembly, disconnect
the ac power cord, remove the four cover screws, and remove the cover. Then proceed to the applicable disassembly
procedure.
Refer to Figure 5-1 for the location of the Electronic Load's mechanical components.

AC Receptacle

1. Record the color code and location of each wire connected to the ac receptacle.
2. Disconnect the push-on connectors from the receptacle terminals.
3. Unsolder the ground wire.
4. Release the locking tabs by pressing them inward against the body of the receptacle and remove the receptacle.

Fan

1. Remove the six screws securing heatsink cover and remove heatsink cover.
2. Disconnect the fan cable from J554.
3. Remove the two screws securing the fan to the main heat sinks and remove the fan.
58
!
turn calibration mode on
!
initialize default calibration parameters
!
store calibration constants in EEROM
!
model number
!
model number suffix and GP-IB address 5
!
initial *SRE value
!
initial *ESE value
!
initial *PSE value
!
module width ( 6063 = 1 CHANNEL )
!
module type
!
voltage for soft over power
!
current for soft over power
!
reset factory default state
!
turn on slew rate
!
to location 0
!
turn calibration mode off