Teledyne 7320 Operating Instructions Manual page 36

Infrared gas analyzer
Table of Contents

Advertisement

3 Start-up and Theory of Operation
Model 7320
Figure II
Linearizer Input
Figure 3-4
Piece-wise approximation is the method used to linearize the signal, i.e., the
linearizer's output to input relationship can be graphed as a number of
straight line segments connected together to approximate the desired curve
that would be required to compensate for the nonlinearity of the bench
The points at which the compensation curve changes slope are called
breakpoints. The slope of each segment corresponds to the gain of the
linearizer in that segment, and this gain has to vary according to the input
voltage. This is achieved through software by adjusting a multiplier factor at
each straight line segment. The linearizer has nine specific voltage segments.
3-8