Observing The Gas Stream Concentrations - Teledyne 4020 Operating & Maintenance Manual

Total hydrocarbon analyzer
Table of Contents

Advertisement

Total Hydrocarbon Analyzer
4.6.16.4 O
BSERVING THE GAS
The actual gas concentrations of the gas streams can be observed
using two different methods:
1. Use the USER SELECT option of the VALVE
SELECTIONS menu and choose the appropriate valve to
open from the second line of the display. Then return to
Analysis mode by repeatedly pressing <Escape> and observe
the concentration of that stream on the display.
Note: With this technique, the alarms are not suppressed and
depending on the gas and the alarm setpoint, it could
trigger an alarm.
2. Use the DEFAULT option of the VALVE SELECTIONS
menu and select the SPAN (or ZERO) function from the
MAIN MENU. With the SPAN (or ZERO) function
selected, select MANUAL FINISH from the first line of the
display on the SPAN or ZERO submenu. The display will
show:
Or
Note: The actual displayed concentration may be different
depending on your calibration gas.
Then begin the span or zero routine and observe the
concentration on the screen. Since the analyzer is in a calibration
mode, the alarms will be suppressed.
Teledyne Analytical Instruments
S
C
TREAM
ONCENTRATIONS
Operation
61

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents