Gas Connections; Installation Checklist - Teledyne 3190 Series Operating Instructions Manual

Trace oxygen analyzer
Hide thumbs Also See for 3190 Series:
Table of Contents

Advertisement

Trace Oxygen Analyzer
is keyed to fit only one way into the receptacle. Do not force it in. The other
end of the cable is made of four separate wires. These should be connected
to the terminal strip on the analysis unit as follows:
Red:
Black:
Green:
White:
Refer to the Final Assembly, Dwg. C-64641.
3.4
Gas connection instructions depend on the specific External Probe used
and any special requirements of the process being monitored.
The standard Model 3190 External Probe has inlet and outlet fixtures
only. Calibration gasses must be tee'd into the sample inlet through appropri-
ate valves. ¼ inch tube fittings are used. For metric installations, ¼ inch to 6
mm adapters are supplied.
In general, sample flow and pressure must not create significant
backpressure past the sensor. For the standard probe, 2 scfh is the nominal
recommended flowrate.
The pressure required will depend on the sampling system. When
venting into a constant pressure, such as the atmosphere, controlling input
pressure is simple. If you are venting into a system of varying pressure, then
some form of pressure regulation is required.
3.5
Before connecting the instrument to the power source and turning it on,
make sure you have:
Once the above checks have been made, you can connect to the power
source. The instrument is now ready for operation.
#1
#2
#3
The green and white connectors can be
}
#4
interchanged, but be consistent.

Gas Connections

Installation Checklist

Correctly installed the Sample and Exhaust gas lines
Opened the isolation valves
Checked for leaks
Set the sample pressure to 5–10 psig, nominal
Teledyne Analytical Instruments
Installation 3
3-7