3. To raise or lower the R-value one step-amount at a time, use the softkeys +R or -R.
4. Continue pressing the softkeys +R or -R until the required R-value displays on the

Adjusting the Step Amount

You can adjust the step to change the value that increases or decreases the R-value with
each button push.
1. First, use the arrow softkey to set the cursor under the step digit that you want to
2. Continue to press --> until the cursor is where you want it. Each time you press -->,
3. With the cursor highlighting the digit you want to change, press +STP to increase the
4. You can now continue to use the softkeys +R or -R to adjust the R-value.
5. When you are done, press esc to return to the Advanced Options menu, then press
11-6
In the sample below, the step amount is 0.020, and thus pressing the softkey +R
raises the previous R-value from 0.610 to 0.630:
LCD.
adjust. The cursor (flashing in the digit field) highlights first "0" in the sample above
of 0.020. Advance to the next digit position by pressing -->.
the cursor moves one position to the right.
numeric value. In the sample that follows, the step is increasing in the second
position by one to 0.120:
esc again to return to Main Menu 2.
esl150.eps
esl151.eps
esl152.eps
esl153.eps