Fluke Biomedical Index 2MF User Manual page 116

Spo2 simulator
Table of Contents

Advertisement

Index 2MF
Users Guide
10 K
20 K
30 K
40 K
Before testing the probe, the Simulator reads this resistor to determine the LED
connection being tested.
Examining the 15-pin connector further:
1. Up to two LED calibration resistors can be tested by the Simulator. Connect them as
2. The probe photodiode is to be connected to pins 9 and 10 as shown.
3. Finally, up to three shields can be tested for shorts to other leads. These would be
Refer to Figure G-3, an example of the Ohmeda probe test connections.
G-4
Table G-1. Resistances with LED Arrangements
Technology Options
R1 and R2 in Figure G-1.
connected to any of pins 11, 12, or 13.
Typical Oximeter
Back to back (typical of Nellcor)
Common cathode
Common Anode (typical of Ohmeda)
Completely separate