Operation - Fluke Biomedical Index 2MF User Manual

Spo2 simulator
Table of Contents

Advertisement

Index 2MF
Users Manual
The Simulator accuracy specifications are not valid for user defined custom
makes. The Simulator can provide specified repeatability of 1 standard
deviation.
Normal Intensity
Low Intensity
High Intensity
Very High Intensity
Curve No.
Once a technology and a curve have been determined, save the custom make for future
use.

Operation

To access the custom menu, press CUST from the main menu.
Chose the slot to customize by pressing +CST to cycle through the available names.
Custom #1 through #5 are the default factory names. To change the name press NAME.
There are five available slots for saving custom makes. Press + or – to change the
alphabet. Press
Press TEC to test for technology type.
Select the appropriate technology type #1 - #4. Technology types offer various light level
thresholds. Refer to Table 5-1 for a list of technology light levels. The technology to use
may already be known based on the manufacturer's information. If not, continue for
process of elimination testing.
Press HiOx to test oximeter stabuility at the high oxygen setting.
5-4
Table 5-1. Oximeters with Technology Options
Technology Options
Table 5-2. Oximeters with R-Curve Parameters
Brand
1
BCI
2
Criticare
3
Datascope
4
Datex
5
Hewlett-Packard
6
Nellcor
7
Hihon Kohden
8
Novametrix
9
Ohmeda
10
Respironics
to move the character. Press SAVE to save the name and the results.
Note
Typical Oximeters
Nellcor, BCI
Hewlett-Packard, Ohmeda, Invivo
Datex
None known
HiOx Level
Oxy @ R = 0.7
93
94
94
95
96
93
94
96
95
93
LoOx Level
Oxy @ R = 1.0
82/83
80/79
85
84
85
80
84
87/88
86/85
80