Changing the Default SpO 2 Setting
The default settings (for the current session from power-up, when no prior changes have
been made) are for a simulated patient with 96% saturated oxygen levels and with a pulse
of 75 beats per minute.
To change either of these settings, press MAN on the Simulations Menu.
The manual settings screen appears on the LCD, displaying the current SpO2 and BPM:
Setting the O 2 Saturation Level
You can raise or lower, as needed, the simulated degree of oxygen saturation in the
virtual patient.
1. To raise the O 2 level two steps at a time, press O2+.
2. To lower the O 2 level two steps at a time, press O2-.
Setting and Changing Simulation Factors
6
Setting Simulations
fgg031.eps
esl032.eps
esl033.eps
esl034.eps
6-3