Resistance ( ) - Fluke 8060A Instruction Manual

True-rms multimeter
Hide thumbs Also See for 8060A:
Table of Contents

Advertisement

8060A
Instruction Manual

2-17. Resistance ( )

Selection of the resistance function is described in Figure 2-12. There are
four fixed ranges (200 , 2 k , 20 k , 200 k ) plus the autoranging M
range consisting of three ranges: 2 M , 20 M , and 300 M .
In all fixed resistance ranges (200 to 200 k ), the test voltage is less than
that required to turn on most semiconductor junctions. This feature,
sometimes referred to as "low power" ohms, aids in troubleshooting by
allowing you to measure resistors independent of the effects of in-circuit
transistors and diodes. For the fixed ranges the maximum full scale voltage
across the circuit being measured is less than 250 mV. The autoranging M
ranges have enough voltage to turn on semiconductor junctions (maximum
2.5V full scale), but the current is very low (2.2 A maximum).

Resistance ( )

1. Select a range
2. Push switch in for
resistance function.
3. Ensure all other switches
are out (except the AC/DC
switch which can be in or out).
4. Connect the test leads as shown.
5. Ensure that the device being measured
contains no electrical energy.
6. Heed the input overload limits (Table 2-2) and
connect the test leads to the device being measured.
7. Read the measured value on the display.
2-18
Figure 2-12. Resistance Operation
Low (-)
High (+)
dx13f.eps