Current Test; Equipment Connection For Current Test - Fluke 8060A Instruction Manual

True-rms multimeter
Hide thumbs Also See for 8060A:
Table of Contents

Advertisement

3.
Apply a resistance of 1.000 M . A reading of 993.0 to 1007.0 should
appear on the display.
4.
Remove the resistance from the inputs to the UUT. The display should
read 0.0 to 0.2.

5-18. Current Test

Use the following procedures to verify proper operation of the dc and ac
current functions:
1.
Connect the UUT and the DMM Calibrator as shown in Figure 5-5.
2.
For each step in Table 5-4, select the UUT function and range as
indicated. Program the DMM Calibrator for the specified input signal
and verify that the displayed UUT value is within the indicated limits.
Figure 5-5. Equipment Connection for Current Test
UUT
Maintenance
Performance Tests
DMM
Calibrator
HI
LO
5
dx36f.eps
5-17