Fluke 8060A Instruction Manual page 63

True-rms multimeter
Hide thumbs Also See for 8060A:
Table of Contents

Advertisement

Operating Instructions
2
Initial Check-Out Procedure
Push the
button twice to enable the visible and audible
continuity. You should see the bar in the display and hear the tone.
Select the diode test (with the V S and A inputs still shorted together).
The display should read .0102 .0015V.
Select the conductance function (with the V S and A inputs still
shorted together). The instrument should indicate overrange (OL).
Remove the connection between the inputs. The instrument should
indicate 0.0 1.0.
2-35