Dc Current Test - Fluke 8842A Instruction Manual

Hide thumbs Also See for 8842A:
Table of Contents

Advertisement

Step
Range
1
20Ω
2
20Ω
3
200Ω
4
200Ω
5
2 kΩ
6
2 kΩ
7
20 kΩ
8
20 kΩ
9
200 kΩ
10
200 kΩ
11
2000 kΩ
12
2000 kΩ
13
20 MΩ
14
20 MΩ
1 – Using offset control
2 = 4 ½ digit counts
3 = Applies to 4-wire ohms only
6-7.

DC Current Test

The following procedure may be used to test the mA DC function:
1. Ensure the 8842A is on and has warmed up for at least 1 hour.
2. Select the mA DC function.
3. Connect the Current Source to the 2A and LO INPUT terminals.
4. For each step in Table 6-6, set the Current Source for the indicated input and verify
that the displayed reading is within the limits shown for each reading rate.
5. Set the Current Source for zero mA and disconnect it from the 8842A.
STEP
RANGE
1
200 mA
2
200 mA
3
2000 mA
4
2000 mA
1000 mA
Table 6-5. Resistance Test
Input
(Nominal)
3
0Ω (short)
3
10Ω
3
0Ω (short)
3
100Ω
0Ω (short)
1 kΩ
0Ω (short)
10 kΩ
0Ω (short)
100 kΩ
0Ω (short)
1 MΩ
0Ω (short)
10 MΩ
Table 6-6. DC Current Test
INPUT
SLOW
MIN
MAX
0 mA
-00.040
+00.040
100 mA
99.920
100.080
0 mA
-000.04
+000.04
999.56
1000.44
1
Error From Input
(in counts)
Slow
Medium
±40
±60
±49
±69
±4
±6
±11
±13
±3
±5
±8
±10
±3
±5
±8
±10
±3
±5
±9
±11
±3
±6
±28
±31
±4
±7
±44
±47
DISPLAYED READING
MEDIUM
MIN
MAX
-00.060
+00.060
99.900
100.100
-000.06
+000.06
999.54
1000.46
Maintenance
PERFORMANCE TEST
2
Fast
±20
±21
±3
±4
±2
±3
±2
±3
±2
±3
±2
±5
±2
±6
FAST
MIN
MAX
-00.20
+00.20
99.76
100.24
-000.2
+000.2
999.4
1000.6
6
6-9