Taking The Measurement In "Auscultation Mode - Omron HBP-1300 Instruction Manual

Professional blood pressure monitor
Hide thumbs Also See for HBP-1300:
Table of Contents

Advertisement

Irregular pulse wave detection function
If the pulse wave interval becomes irregular during measurement, the irregular pulse wave detection
icon will light to notify you.
Body movement detection function
If body movement is detected during measurement, deflation stops for 5 seconds. The irregular pulse
wave icon appears in the measurement result display.
■ Deflation stopped
After 5 seconds, measurement resumes, and an attempt is made to complete measurement in one
cycle.

Taking the Measurement in "Auscultation Mode"

Use a stethoscope to perform measurement.
When set to "ON", SYS and DIA can be recorded using auscultation measurement.
To determine SYS and DIA during measurement, press the [MODE] button.
SYS is registered the first time and DIA is registered the second time you press the [MODE] button
during deflation.
After DIA is determined, the cuff rapidly deflates and SYS and DIA are displayed as the measurement
results.
The pulse rate does not appear in the measurement result display.
During deflation, the [
deflate faster.
Stored data that was measured in "Auscultation Mode" is displayed when the auscultation icon is lit.
■ During re-inflation
Note:
The body movement detection function is disabled while "Auscultation Mode" is in use.
For Auscultation Mode settings, see page 16.
22
] button can be held down to re-inflate, or the [
] button can be held down to

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents