Command 4 - Conversion Equation Setup - Casio CFX-9850G User Manual

Graphic scientific calculator
Hide thumbs Also See for CFX-9850G:
Table of Contents

Advertisement

Chapter 3: Commands and Programming
SAMPLING/TRIGGER SETUP Example
{3,0.2,50}
{3, 0.001, 255, 1, 2, 2}

Command 4 - CONVERSION EQUATION SETUP

Command 4 specifies conversion type, order, and constants for conversion
to special units (like Newtons or degrees Celsius).
Syntax
{4, Equation Number, Equation Type, Units Display, Constants}
Equation Number Parameter
This parameter specifies a equation number for loading of conversion equa-
tions. Equations have a fixed relationship with channels, so that Equation 1
corresponds to Channel 1, Equation 2 corresponds to Channel 2, and so on.
The initial default for this parameter is 0.
50
0
No filtering
1
5-point S-G smoothing
2
9-point S-G smoothing
3
17-point S-G smoothing
4
25-point S-G smoothing
5
3-point median filtering
6
5-point median filtering
• No filtering is performed when the number of samples parameter of Com-
mand 3 is –1 (real-time sampling).
0
Clears all equations.*
1
Equation 1 (Channel 1)
2
Equation 2 (Channel 2)
3
Equation 3 (Channel 3)
4
Equation 4 (SONIC channel)
3 = SAMPLING/TRIGGER SETUP
0.2 = Sampling every 0.2 second
50 = 50 samples
Initial defaults are used for triggering and other
parameters.
3 = SAMPLING/TRIGGER SETUP
0.001 = Sampling every 1 millisecond
255 = 255 samples
1 = Absolute time recording
2 = External trigger (CH1)
2 = Trigger on voltage greater than 2V

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cfx-9800gEa-100

Table of Contents