Introduction:
Introduction:
Introduction:
Introduction:
Introduction:
The Hanna HI 4003 and HI 4103 are ion selective elec-
trodes designed for the measurement of cadmium ions in
aqueous solutions. The HI 4003 is a solid state half-cell
sensor that requires a separate reference. The HI 4103 is a
combination ion selective electrode.
II II II II II. Specifications
Specifications
Specifications
Specifications
Specifications
Type:
Ion(s) measured:
Measurement range:
Interfering ions:
be absent. Any ion that forms a more insoluble sulfide than
cadmium sulfide will interfere with the electrode response.
2+
These include lead (Pb
)and cupric ions (Cu
of the cadmium surface by ferric ion( Fe
poison the electrode. Ions that can form complexes with
cadmium ions will reduce the measured concentration.
Examples include bromide, chloride and EDTA.
Operating Temperature:
Operating pH:
Dimensions:
Connection:
Solid State electrode with
a Cadmium sulfide/ silver
sulfide membrane.
2+ + + + +
Cadmium (Cd
)
0.1 M to 1X 10
-7
M
11,200 to 0.01 ppm
Silver and mercury must
2+
). Oxidation
3+
) will severely
0-80°C
2 to12 pH
concentration dependant
(4.5-7.5 recommended)
12 mm (OD) X 120 mm
nominal insertion
(0.47" X 4.72")
BNC
2