Hanna Instruments HI 8915 Instruction Manual
Hanna Instruments HI 8915 Instruction Manual

Hanna Instruments HI 8915 Instruction Manual

Portable ph meters
Hide thumbs Also See for HI 8915:

Advertisement

Instruction Manual
HI 8014 • HI 8915
HI 931000
Portable pH Meters
w w w . h a n n a i n s t . c o m
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HI 8915 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hanna Instruments HI 8915

  • Page 1 Instruction Manual HI 8014 • HI 8915 HI 931000 Portable pH Meters w w w . h a n n a i n s t . c o m...
  • Page 2: Table Of Contents

    If the repair is not covered by the warranty, you will be notified of GENERAL DESCRIPTION the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid.
  • Page 3: Functional Description Hi8014

    FUNCTIONAL DESCRIPTION HI8014 SPECIFICATIONS HI8014 HI 8014 Range 0.00 to 14.00 ±1999 Resolution 0.01 pH meter Accuracy ±0.01 HI 8014 (@20°C/68°F) mV ±1 °C Typical EMC ±0.05 Deviation ±8 pH Calibration Manual 2 point through offset and slope trimmers Offset Calibration ±1 pH Slope Calibration from 85 to 105%...
  • Page 4: Functional Description Hi8915

    FUNCTIONAL DESCRIPTION HI8915 SPECIFICATIONS HI8915 HI 8915 Range 0.00 to 14.00 ±1999 °C 0.0 to 100.0 Resolution 0.01 °C Accuracy ±0.01 HI 8915 (@20°C/68°F) mV ±1 pHmeter °C ±0.5 HOLD Typical EMC ±0.02 Deviation ±1 °C °C ±0.8 OFFSET SLOPE pH Calibration Manual 2 point through 6.86...
  • Page 5: Functional Description Hi931000

    FUNCTIONAL DESCRIPTION HI931000 SPECIFICATIONS HI931000 HI 931000 Range 0.00 to 14.00 °C 0.0 to 70.0 ins rum n s Resolution 0.01 °C Accuracy ±0.01 (@20°C/68°F) °C ±0.5 Typical EMC ±0.30 Deviation °C ±1 °C pH Calibration Automatic 2 point with 3 memorized standard buffers (pH 7.01, 4.01, 10.01) HI 931000...
  • Page 6: Operational Guide

    pH MEASUREMENTS OPERATIONAL GUIDE For HI 8014: To take a pH measurement simply submerge INITIAL PREPARATION the tip (4cm/1½") of the electrode and a Each meter is supplied complete with a 9V battery. Slide off the ChecktempC (or another accurate thermom- battery compartment cover on the back of the meter, install the eter) in the sample to be tested.
  • Page 7 Notes: Dip the temperature probe or the pH/°C electrode into the sample and allow the reading to stabilize (a couple of minutes). • In order to take accurate pH measurements, make sure that the instrument is calibrated for pH before use. For HI 931000, to display the measured tem- perature, press and hold the °C key.
  • Page 8: Ph Calibration

    • Set the rotary knob to °C position to display the manual pH CALIBRATION temperature setting and adjust the °C trimmer until the LCD shows the measured temperature. For best accuracy, frequent calibration of the instrument is recom- °C mended. The instrument should be recalibrated for pH: a) Whenever the pH electrode or temperature probe is replaced.
  • Page 9 • Stir gently and wait a couple of minutes for thermal equilibrium. PROCEDURE FOR HI 931000 Note: The electrode should be submerged approximately 4 cm (1½") • Turn the meter on. into the solution. The temperature probe should be located as Note: The "...
  • Page 10: Ph Values At Various Temperatures

    pH VALUES AT VARIOUS TEMPERATURE CALIBRATION TEMPERATURES HI 8915 and HI 931000 have been accurately precalibrated for For manual temperature compensation during calibration, please temperature at the factory. refer to the following chart. For optimum accuracy, it is recommended that you have your pH meter recalibrated for temperature at least once a year.
  • Page 11: Electrode Conditioning And Maintenance

    If the bulb and/or junction are dry, soak the electrode in HI 70300 ELECTRODE CONDITIONING AND Storage Solution for at least one hour. MAINTENANCE For refillable electrodes: If the refill solution (electrolyte) is more than 2½ cm (1") below the fill hole, add HI 7082 (3.5M KCl Electrolyte Solution) for double junction or HI 7071 (3.5M KCl+AgCl Electrolyte Solution) for single junction electrodes.
  • Page 12: Battery Replacement

    Inorganic Soak in Hanna HI 7074 Inorganic Cleaning Solu- BATTERY REPLACEMENT tion for 15 minutes. Oil /grease Rinse with Hanna HI 7077 Oil and Fat Cleaning All meters are powered by a 9V battery that is located on the rear of Solution.
  • Page 13: Accessories

    ORP SOLUTIONS ACCESSORIES HI 7020M Test solution, 200-275 mV, 230 mL bottle HI 7020L Test solution, 200-275 mV, 500 mL bottle pH CALIBRATION SOLUTIONS HI 7091M Reducing pretreatment solution, 230 mL bottle HI 70004P pH 4.01 buffer solution, 20 mL sachet, 25 pcs HI 7091L Reducing pretreatment solution, 500 mL bottle HI 7004M...
  • Page 14: Ce Declaration Of Conformity

    FC210B Glass body, double junction, conic, Viscolene, non- CE DECLARATION OF CONFORMITY refillable combination pH electrode. Use: milk, yogurt. FC220B Glass body, triple ceramic, single junction, refill- able, combination pH electrode. Use: food processing. FC911B PVDF body, double junction, refillable with built- in amplifier, combination pH electrode.
  • Page 15 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel.