FUNCTIONAL DESCRIPTION
WARRANTY
This meter is warranted for two years against defects in workmanship and materials
when used for their intended purpose and maintained according to instructions. Elec-
trodes are warranted for six months. This warranty is limited to repair or replacement
free of charge. Damage due to accidents, misuse, tampering or lack of prescribed main-
tenance are not covered. If service is required, contact the dealer from whom you pur-
chased the instrument. If under warranty, report the model number, date of purchase,
serial number and the nature of the problem. If the repair is not covered by the war-
ranty, you will be notified of the charges incurred. If the instrument is to be returned to
Hanna Instruments, first obtain a Returned Goods Authorization Number from the Cus-
tomer Service department and then send it with shipment costs prepaid. When shipping
any instrument, make sure it is properly packaged for complete protection.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material and examine it
carefully to make sure that no damage has occurred during ship-
ment. If noticeable damage is found, notify your dealer.
HI 8010 comes supplied complete with HI 1230B pH electrode, cali-
bration screwdriver, 1 x 9V battery and instructions.
Note: Save all packing material until you are sure that the instru-
ment functions correctly. Any defective items must be returned
in the original packing together with the supplied accessories.
SPECIFICATIONS
Range
0.00 to 14.00 pH
Resolution
0.01 pH
Accuracy
±0.01 pH
(@20°C/68°F)
Typical EMC Deviation
±0.05 pH
Calibration
Manual, 2-point through trimmers
Temperature
Manual, through trimmer
Compensation
from 0 to 100°C (32 to 212°F)
Probe
HI 1230B double junction, gel-filled,
pH electrode with BNC connector and 1 m cable (included)
Environment
0 to 50°C (32 to 122°F); RH max 95%
Power Supply
1 x 9V battery
approx. 100 hours of continuous use
Dimensions
185x82x53 mm (7.3x3.2x2.1'')
Weight
265 g (9.3 oz.)
HI 8010 is in compliance with the
directives.

Advertisement

loading
Need help?

Need help?

Do you have a question about the HI 8010 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hanna Instruments HI 8010

  • Page 1 If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Cus- tomer Service department and then send it with shipment costs prepaid. When shipping w w w .
  • Page 2 OPERATIONAL GUIDE CALIBRATION PROCEDURE • Rinse and immerse pH electrode and thermometer in pH 4.01 or pH 10.01 buffer (2 calibration point); stir gently. • Slide off the battery compartment cover on the back of the meter For better accuracy, a frequent calibration of the meter is recommended. •...