Fluke 374 Calibration Manual page 20

Clamp meter
Hide thumbs Also See for 374:
Table of Contents

Advertisement

374/375/376
Calibration Manual
The calibration adjustment is complete.
When the adjustment is complete:
1. Remove the Power Supply.
2. Replace the batteries.
3. Reattach the battery compartment door.
4. Tighten the battery compartment door screw.
12
Figure 4. Current Calibration Setup
5522A
CALIBR ATOR
ghn50.eps