Fluke 123 User Manual page 80

Industrial scopemeter
Hide thumbs Also See for 123:
Table of Contents

Advertisement

Fluke 123/124
Users Manual
The Fluke 123/124, including standard accessories, conforms with the EEC directive 2004/108/EC for EMC immunity, as
defined by EN-61326-1: 2006, with the addition of the following tables.
Trace disturbance with STL120
No visible disturbance
Frequency range 10 kHz to 27 MHz
Frequency range 27 MHz to 1 GHz
No visible disturbance
Frequency range 1.4 GHz to 2.0 GHz
Frequency range 2.0 GHz to 2.7 GHz
Disturbance less than 10% of full scale
Frequency range 10 kHz to 27 MHz
Frequency range 27 MHz to 1 GHz
(-): no visible disturbanceTest tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full
scale.
72
Table 1
E= 3 V/m
100 mV/div to 500 V/div
100 mV/div to 500 V/div
E= 3 V/m
5 mV/div to 500 V/div
Table 2
E= 3 V/m
20 mV/div to 50 mV/div
10 mV/div to 20 mV/div
E= 10 V/m
500 mV/div to 500 V/div
100 mV/div to 500 V/div
E= 1 V/m
5 mV/div to 500 V/div
E= 10 V/m
100 mV/div to 200 mV/div
-

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

124124/s123/s

Table of Contents