Important
Please read this "Instructions for Use", and "Installation Manual"
(separate volume) carefully to familiarize yourself with safe and
effective usage.
Please retain this manual for future reference.
• For monitor adjustment and settings, refer to the "Installation
Manual".
• For the latest product information including the "Instructions for
Use", refer to our web site :
http://www.eizoglobal.com

Advertisement

Table of Contents
loading

Summary of Contents for Eizo RadiForce RX360

  • Page 1 Instructions for Use Color LCD Monitor Important Please read this “Instructions for Use”, and “Installation Manual” (separate volume) carefully to familiarize yourself with safe and effective usage. Please retain this manual for future reference. • For monitor adjustment and settings, refer to the “Installation Manual”.
  • Page 2 EIZO Corporation is under no obligation to hold any submitted material or information confidential unless prior arrangements are made pursuant to EIZO Corporation’s receipt of said information. Although every effort has been made to ensure that this manual provides up-to-date information, please note that EIZO monitor specifications are subject to change without notice.
  • Page 3: Precautions

    PRECAUTIONS IMPORTANT • This product has been adjusted specifically for use in the region to which it was originally shipped. If the product is used outside the region, it may not operate as specified in the specifications. • To personal safety and proper maintenance, please read carefully this section and the caution statements on the monitor.
  • Page 4 A unit placed on an inadequate surface may fall and result in injury or equipment damage. If the unit falls, disconnect the power immediately and ask your local EIZO representative for advice. Do not continue using a damaged unit. Using a damaged unit may result in fire or electric shock.
  • Page 5 If the unit falls, please ask your local EIZO representative for advice. Do not continue using a damaged unit. Using a damaged unit may result in fire or electric shock. When reattaching the tilt stand, please use the same screws and tighten them securely.
  • Page 6 CAUTION Handle with care when carrying the unit. Disconnect the power cord and cables when moving the unit. Moving the unit with the power cord or cables attached is dangerous and may result in injury. Carry or place the unit according to the correct specified methods. •...
  • Page 7: Notice For This Monitor

    When the screen becomes dark or begins to flicker, please contact your local EIZO representative. • The screen may have defective pixels or a small number of light dots on the screen. This is due to the characteristics of the panel itself, and is not a malfunction of the product.
  • Page 8: To Use The Monitor For A Long Time

    • To adjust measurement results of the integrated calibration sensor (Integrated Front Sensor) to those of an EIZO external sensor (UX1 or UX2 sensor) that is sold separately, perform correlation between the Integrated Front Sensor and the external sensor using RadiCS / RadiCS LE. Periodical correlation allows you to maintain the measurement accuracy of the Integrated Front Sensor at a level equivalent to that of the external sensor.
  • Page 9: Table Of Contents

    CONTENTS ............. 9 Chapter 1 Introduction ........10 1-1. Features ............10 1-2. Package Contents ........11 ● EIZO LCD Utility Disk ........11 1-3. Controls and Functions ......12 Chapter 2 Installation / Connection ....13 2-1. Before Installing the Product .....13 ●...
  • Page 10: Chapter 1 Introduction

    Chapter 1 Introduction Thank you very much for choosing an EIZO color LCD monitor. 1-1. Features ● Hybrid monochrome and color display When the Hybrid Gamma PXL function is enabled, this product automatically differentiates between monochrome and color parts of the same image at a pixel level, and displays them respectively in optimal gradations.
  • Page 11: Package Contents

    1-2. Package Contents Check that all of the following items are contained in the package. If any of these are missing or damaged, contact your dealer or local EIZO representative. Note • It is recommended that the box and packing materials be stored so that they can be used to move or transport this product.
  • Page 12: Controls And Functions

    1-3. Controls and Functions Approx. 30° 1. Integrated Front Sensor This sensor is used to perform calibration and Grayscale Check. (Movable) 2. Ambient Light Sensor This sensor measures environmental illumination. Environmental illuminance measurement is performed using the RadiCS / RadiCS LE quality control software. 3.
  • Page 13: Chapter 2 Installation / Connection

    Chapter 2 Installation / Connection 2-1. Before Installing the Product Carefully read “PRECAUTIONS” (page 3) and always follow the instructions. If you place this product on a lacquer-coated desk, the color may adhere to the bottom of the stand due to the composition of the rubber.
  • Page 14: Connecting Cables

    2-2. Connecting Cables Attention • Check that the monitor and the PC are powered off. • When replacing the current monitor with this monitor, refer to “4-2. Compatible Resolutions” (page 20) to change the PC settings for resolution and vertical scan frequency to those that are available for this monitor, before connecting the PC.
  • Page 15 Attention • Visit the EIZO website for information about monitors and graphic boards that can be used for the daisy- chain connection: http://www.eizoglobal.com • When using a daisy-chain connection, connect the cable to the input connector •...
  • Page 16 When you use RadiCS / RadiCS LE or connect a USB device (peripheral device that supports USB) to the monitor, connect the USB cable to the monitor's USB upstream port and the PC. USB cable: UU300 Attention • When connecting the monitor to a PC on which RadiCS / RadiCS LE has been installed, connect the cable •...
  • Page 17: Turning On The Power

    2-3. Turning On the Power Touch to turn on the power to the monitor. The power switch indicator of the monitor lights up green. If the indicator does not light up, see “Chapter 3 No-Picture Problem” (page 18). Note • To find the location of the power switch when the monitor power is shut off, touch any of the buttons besides to make the indicator flash.
  • Page 18: Chapter 3 No-Picture Problem

    • Turn off the main power, and then turn it on again. • Power switch indicator blinks: Orange, • Connect via the signal cable specified by EIZO. Turn off the Green main power, and then turn it on again.
  • Page 19: Chapter 4 Specifications

    Chapter 4 Specifications 4-1. Specifications List Type RX360: Anti-Glare RX360-AR: Anti-Reflection LCD Panel Type Color (IPS) Backlight Size 54.1 cm (21.3 inch) Resolution (H x V) 1536 × 2048 Display Size (H x V) 324.9 mm × 433.2 mm Pixel Pitch 0.2115 mm Display Colors 10-bit (DisplayPort): 1073.74 million colors (Max.) (from a palette...
  • Page 20: Compatible Resolutions

    Transportation Temperature -20 ˚C - 60 ˚C (-4 ˚F - 140 ˚F) / Storage Humidity 10 % - 90 % R.H. (no condensation) Environmental Air Pressure 200 hPa - 1060 hPa Requirements *1 When the PinP sub window is displayed *2 The vertical scan frequency supported varies according to the resolution.
  • Page 21: Appendix

    Appendix Medical Standard • It shall be assured that the final system is in compliance to IEC60601-1-1 requirement. • Power supplied equipment can emit electromagnetic waves, that could influence, limit or result in malfunction of the monitor. Install the equipment in a controlled environment, where such effects are avoided.
  • Page 22: Emc Information

    Be sure to use the cables attached to the product, or cables specified by EIZO. Use of cables other than those specified or provided by EIZO of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and improper operation.
  • Page 23 Technical Descriptions Electromagnetic emissions The RadiForce series is intended for use in the electromagnetic environment specified below. The customer or the user of the RadiForce series should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment - Guidance RF emissions Group 1 The RadiForce series uses RF energy only for its internal function.
  • Page 24 Electromagnetic immunity The RadiForce series has been tested at the following compliance levels according to the testing requirements for professional healthcare facility environments defined in IEC / EN60601-1-2. Customers and users of the RadiForce series must ensure that the RadiForce series is used in the following environments: Immunity test Test level for...
  • Page 25 Recommended separation distances between portable or mobile RF communication equipment and the RadiForce Series The RadiForce series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the RadiForce series can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the RadiForce series.
  • Page 26 The RadiForce series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. For other portable and mobile RF communication equipments (transmitters), minimum distance between portable and mobile RF communications equipment (transmitters) and the RadiForce series as recommended below, according to the maximum output power of the communications equipment.
  • Page 27: Fcc Declaration Of Conformity

    Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. Note Use the attached specified cable below or EIZO signal cable with this monitor so as to keep interference within the limits of a Class B digital device. - AC Cord...
  • Page 28 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan 中国苏州市苏州工业园区展业路 8 号中新科技工业坊 5B Siemensallee 84, 76187 Karlsruhe, Germany 00N0N098B1 http://www.eizoglobal.com IFU-RX360-6 Copyright © 2018 EIZO Corporation. All rights reserved. 2nd Edition-November, 2018 Printed in Japan.