Lifescan BASIC - 11-96 Owner's Booklet page 50

Table of Contents

Advertisement

MESSAGE
c 68
(example)
contrl
error 1
retest
PROBLEM
4) The center of the Test
Strip Holder where the
small hole is located is
raised.
If this message appears after
a blood test, your blood
sample was too small,
smeared, or another drop was
added after the test began.
The sample was applied
before apply sample
appeared.
WHAT TO DO
Depress the center by
pressing forward on the
base, or by pressing
down on the center, or by
inserting a Check Strip
into the Test Strip Holder.
Repeat the test with a
new Test Strip and
a larger drop of blood.
Repeat test with a new
Test Strip and wait for
apply sample to appear.