Summary of Contents for EUTECH INSTRUMENTS CYBERSCAN PH 5000 PH METER
Page 1
Instruction Manual CyberScan pH 5500/ 5000 Bench pH & Ion Meter/ Bench pH Meter 68X292328 Rev. 1 07/03 Technology Made Easy ...
Page 3
Eutech Instruments’ Authorized Distributors or call us at (65) 6778-6876 for Eutech Instruments’ Customer Service Dept. for assistance. Kindly remember to complete the warranty card and mail it back to your Authorized Distributors or Eutech Instruments Pte Ltd.
Instruction Manual CyberScan pH5500/ 5000 TABLE OF CONTENTS INTRODUCTION Introducing the CyberScan Series 5000 UNPACKING THE METER GETTING STARTED Connectors pH Electrodes & Ion Specific Electrodes USING THE METER Touch screen Operation Choosing a Channel (For CyberScan pH5500 only) Button Functions SYSTEM SETUP Access System Setup Select Language...
Page 5
Instruction Manual CyberScan pH5500/ 5000 11 ION OPERATION 11.1 Direct Reading Methods 11.2 Incremental Methods 11.3 Known Addition Method 11.4 Known Subtraction Method 11.5 Analate addition 11.6 Analate Subtraction 12 CLEANING 13 TROUBLE SHOOTING 14 COMPLIANCE 15 APPENDIX 15.1 Determining Isopotential Points Experimentally 15.2 Data Management 15.3...
CyberScan pH5500/ 5000 INTRODUCTION 1.1 Introducing the CyberScan Series 5000 Thank you for selecting a Eutech Instruments CyberScan Bench meter. This manual describes the operation of the CyberScan Series 5000 bench meters. The state-of-art meter that you have purchased is easy to operate and will guide you through the various functions by displaying easy to understand prompts.
ATC probe Instruction Manual If any of these items are missing, please contact your nearest Authorized Distributor. Accessory conductivity probes and ion selective electrodes are available and can be ordered by contacting any Authorized Distributor or Eutech Instruments Marketing (65) 67786876.
Instruction Manual CyberScan pH5500/ 5000 GETTING STARTED 3.1 Connectors 1. Review the layout and arrangement of the rear connector panel.
Page 9
Instruction Manual CyberScan pH5500/ 5000 2. Connect the electrode arm to the base. 3. Connect the power cable to the connector cable to the rear connector panel power jack and to a power source. To connect RS232, see Data Management on Section 15.2.
Instruction Manual CyberScan pH5500/ 5000 3.2 pH Electrodes & Ion Specific Electrodes (Ion Specific Electrode applicable for CyberScan pH5500 only) This meter allows you to use two types of pH electrodes: the conventional glass pH electrode and the field effect transistor (FET) pH electrode.
Page 11
Instruction Manual CyberScan pH5500/ 5000 Option: Connect the optional FET electrode by plugging it into the FET jack on the back meter panel. Allow the FET electrode to warm up five minutes before use. Connect ion specific electrodes in the same manner as pH electrodes.
Instruction Manual CyberScan pH5500/ 5000 4.1 Touch screen Operation This new CyberScan research benchtop meter operates with a state of art touch screen. The touch screen makes this the easiest meter on the market to operate and care for. When this meter is first plugged in, the STANDBY screen will appear.
Instruction Manual CyberScan pH5500/ 5000 4.2 Choosing a Channel (For CyberScan pH5500 only) Main Screen Channel Screen E U T E C H E U T E C H INSTRUMENTS INSTRUMENTS CyberScan pH 5500 CyberScan pH 5500 dual Select from the options to the right Select from the options to the right setup January 15, 2002...
Page 15
Instruction Manual CyberScan pH5500/ 5000 The CyberScan pH5500 is a multi-channel meter. With this meter you can switch from channel 1 to channel 2. You can also view both channels at the same time by accessing the dual channel mode. If you set the meter to view both channels, you cannot change the parameters without choosing the screen that you want to modify.
Instruction Manual CyberScan pH5500/ 5000 4.3 Button Functions The touch screen of your CyberScan bench meter has “buttons” along the right side of the screen that are common to many of the screens. The following indicates the function of these common buttons.
Page 17
Instruction Manual CyberScan pH5500/ 5000 This button accesses the standardization screen from the various measurement modes and initiates standardization of the meter once the standardization screen is accessed. This button is the measure button and directs the meter to meas measure your sample when in the Auto Read function of the pH mode.
Instruction Manual CyberScan pH5500/ 5000 SYSTEM SETUP System Setup SYSTEM SETUP OPTIONS Select Language - Set Date - Set Time - Set Beeper Status - Set Print Configuration edit - Set Operator - Set Display Contrast - Display Meter Information - Reset to Factory Defaults...
Instruction Manual CyberScan pH5500/ 5000 5.1 Access System Setup E U T E C H E U T E C H INSTRUMENTS INSTRUMENTS CyberScan pH 5500 CyberScan pH 5500 dual Select from the options to the right Select from the options to the right setup January 15, 2002 11:11 am...
Page 20
Instruction Manual CyberScan pH5500/ 5000 The system setup function allows you to customise the meter display options to meet your personal preference. Once set, these will rarely need to be changed. To access System Setup Touch anywhere on the standby screen Touch setup on the main screen Touch system on the setup screen The system setup options are now displayed on the screen.
Instruction Manual CyberScan pH5500/ 5000 5.2 Select Language Set Language English Current LANGUAGE English French German Spanish enter arrow keys to highlight desired language and then touch enter to accept...
Page 22
Instruction Manual CyberScan pH5500/ 5000 This option allows you to choose the language in which all prompts and directions will appear on the touch screen. To Select Language 1. Access the select language screen from the system setup screen. The current language is displayed on the screen. 2.
Instruction Manual CyberScan pH5500/ 5000 5.3 Set Date Set Date enter Current MM/DD/YY D/M/Y clear Touch clear to delete current date Use numeric touchpad to input the new date and then touch enter to accept...
Page 24
Instruction Manual CyberScan pH5500/ 5000 This screen can be used to set the present date which will be displayed on the measure screens. This date will also be printed on demand and stored in the data storage center of the meter when data is saved.
Instruction Manual CyberScan pH5500/ 5000 5.4 Set Time Set Time enter Current 12HR: 10:40 am 24hr clear Touch clear to delete current time numeric touchpad to input the new time and then touch enter to accept...
Page 26
Instruction Manual CyberScan pH5500/ 5000 This screen can be used to set the present time which will be displayed on the measure screens. This time will also be printed on demand and stored in the data storage center of the meter when data is saved.
Instruction Manual CyberScan pH5500/ 5000 5.5 Set Beeper Status Set Beeper Status Current BEEPER STATUS After STABLE Reading - After KEY Touch - On LIMIT Exceeded save arrow keys to highlight beeper option and then touch ON/OFF change Touch save to accept changes...
Page 28
Instruction Manual CyberScan pH5500/ 5000 This screen allows you to turn on or off the beeper. You may choose to have an audible signal when the meter recognizes that the current measurement is stable, each time a function button is touched and/ or when the set limits of a measurement mode have been exceeded.
Instruction Manual CyberScan pH5500/ 5000 5.6 Select Procedural Level Select Procedural Level enter Current LEVEL LEVEL 1 - Basic Level LEVEL 2 - Advanced Level Procedural Level 1 January 15, 2002 11:11am Temperature 25.0°C Electrode Performance 100% Procedural Level 2 January 15, 2002 11:11am 00000...
Page 30
Instruction Manual CyberScan pH5500/ 5000 This selection screen allows you to choose the amount of information that you want to have displayed on the screen. There are two levels to choose from. Both of the levels provide identical results. The amount of information appearing on the measure screens and the number of setup parameters you can manipulate will vary from Basic to Advanced procedural levels.
Instruction Manual CyberScan pH5500/ 5000 5.7 Set Print Configuration Set Print Configuration Current PRINT CONFIGURATION - Baud Rate 19200 - Number of Bits - Stop Bits - Parity NONE - Serial Output Mode TEXT edit arrow keys to highlight desired print setting and then touch edit change...
Page 32
Instruction Manual CyberScan pH5500/ 5000 You can adjust the print configuration of this meter from this screen. The configuration of the following screens must match the configuration of the printer or computer to which the data will be sent. To Set Print Configuration 1.
Instruction Manual CyberScan pH5500/ 5000 5.8 Set Baud Rate This configuration option will control the speed at which the data will be transmitted by the printer. This parameter needs to match the baud rate designated by the printer or computer. To Set Baud Rate 1.
Instruction Manual CyberScan pH5500/ 5000 5.9 Set Number of Bits To Set Number of Bits 1. Access the Set number of Bits screen from the Set Print Configuration screen. The current number of bits is displayed on the screen. 2. Touch 7 or 8 to select the number of bits. 3.
Instruction Manual CyberScan pH5500/ 5000 5.10 Set Stop Bits To Set Stop Bits 1. Access the Set Stop Bits screen from the Set Print Configuration screen. The current number of bits is displayed on the screen. 2. Touch 1 or 2 to set the desired number of stop bits. 3.
Instruction Manual CyberScan pH5500/ 5000 5.11 Set Parity To Set Parity 1. Access the Set Parity screen from the Set Print Configuration screen. The current Parity is displayed on the screen. 2. Touch ODD or EVEN or NONE to set the desired parity. 3.
Instruction Manual CyberScan pH5500/ 5000 5.12 Set Operator Set Operator enter Current Operator clear...
Page 38
Instruction Manual CyberScan pH5500/ 5000 This option allows you to identify the user of the meter. This information can be saved in the meter’s memory. It can also be printed out with measurement data on demand. The operator identification can be up to 9 characters in length. To Set Operator 1.
Instruction Manual CyberScan pH5500/ 5000 5.13 Set Display Contrast Set Display Contrast lighter darker save Touch lighter darker to adjust contrast and then touch save to accept The value displayed ranges from 0(darkest) to 25 (lightest)
Page 40
Instruction Manual CyberScan pH5500/ 5000 This option allows you to change the contrast on the screen to improve the readability of the information presented on the screen. The numbering system that appears on the screen is from 0 to 25. The darkest setting is 0 and the lightest setting is 25.
Instruction Manual CyberScan pH5500/ 5000 5.14 Display Meter Information This screen displays the model number, serial number and current software revision of your meter. E U T E C H INSTRUMENTS CyberScan pH 5500 Unit Serial Number: XX999XXX Software Revision: 1.04b...
Instruction Manual CyberScan pH5500/ 5000 5.15 Reset to Factory Defaults This screen allows you to reset all functions and setup parameters of the meter of the settings originally programmed at the factory. To Reset to Factory Defaults 1. Access the Reset to Factory Defaults screen from the System Setup screen.
Instruction Manual CyberScan pH5500/ 5000 6.1 Access pH setup E U T E C H Setup INSTRUMENTS CyberScan pH 5500 Select from the options to the right Select from the options to the right setup system Last std: Jan 15 @ 11:11 am pH Setup There are two ways to access the pH SETUP OPTIONS...
Page 45
Instruction Manual CyberScan pH5500/ 5000 Measure E U T E C H INSTRUMENTS channel 1 7.00 meas CyberScan pH 5500 BUFFERS Last std: Jan 15 @ 11:11 am Touch meas to measure sample Select from the options to the right Touch to access standardize setup...
Instruction Manual CyberScan pH5500/ 5000 6.2 Set Sample ID# Set pH Sample ID# enter Manual ID# clear...
Page 47
Instruction Manual CyberScan pH5500/ 5000 When this option is active, each time you touch print on the Measure screen the pH value along with the date/ time/ channel and the sample ID# will be sent to data storage. (See Data Storage Criteria page 68 for additional information on saved parameters).
Page 48
Instruction Manual CyberScan pH5500/ 5000 Set pH Sample ID# enter Sequential ID# 24hr clear Touch clear to delete current ID numeric touchpad to input the new starting ID and then touch enter to accept...
Page 49
Instruction Manual CyberScan pH5500/ 5000 Sequential ID# Assignment 1. Access the Set Sample ID# screen from the pH (mV, Ion) Setup screen. 2. Touch seq for sequential ID# assignment. The current ID# is displayed on the screen. 3. Touch clear to delete the current ID#. 4.
Instruction Manual CyberScan pH5500/ 5000 6.3 Select Buffer Group View pH Buffer Group enter Current Buffer Group EURO NIST custom keys to display desired buffer group and then touch enter to accept...
Page 51
Instruction Manual CyberScan pH5500/ 5000 This setup option allows you to select from 3 different buffer groups each containing 5 buffers, for auto buffer recognition. Or you can create a custom group of buffers for auto buffer recognition by touching custom. The 3 existing buffer groups are: USA buffers: 2,4,7,10 and 12...
Instruction Manual CyberScan pH5500/ 5000 6.4 Set pH Custom Buffer Group View pH Custom Buffer Group View pH Custom Buffer Group enter Current BUFFER GROUP Current BUFFER GROUP CUSTOM CUSTOM edit edit enter keys to display desired buffer group keys to display desired buffer group and then touch enter...
Page 53
Instruction Manual CyberScan pH5500/ 5000 To Set pH Custom Buffer Group This option allows you to create a custom buffer group of up to 5 buffers to be used for auto buffer recognition. To obtain optimal results, it is important to maintain at least 2 pH units between selected buffers in the custom group.
Instruction Manual CyberScan pH5500/ 5000 Select Buffer Recognition This option allows you to select automatic buffer recognition or manual buffer recognition when standardizing. With the Automatic buffer recognition activated, the meter will automatically recognize the buffers from the chosen buffer group and accept them when the meter recognizes the reading as stable.
Instruction Manual CyberScan pH5500/ 5000 6.6 Select Auto Read Mode You can use this meter when the Auto Read Mode function is active or when it is inactive. When the Auto Read function is active, the meter will lock onto a reading when the meter recognizes it as stable. The meter will not deviate from this reading until meas is touched.
Instruction Manual CyberScan pH5500/ 5000 6.7 Set pH Stability Criteria Set pH Stability Criteria Medium Current CRITERIA Fast Medium Slow enter arrow keys to highlight stability criteria and then touch enter to accept...
Page 57
Instruction Manual CyberScan pH5500/ 5000 This setup screen allows you to determine how quickly the meter will respond to electrode drift. There are 3 speed settings: fast, medium and slow. To set pH Stability Criteria 1. Access the Set pH Stability Criteria screen from the pH (Ion) Setup screen.
Instruction Manual CyberScan pH5500/ 5000 6.8 Set Default Temperature Set pH Default Temperature enter 25.0°C Current DEFAULT clear Touch clear to delete current default temp numeric touchpad to input the new temperature and then touch enter to accept...
Page 59
Instruction Manual CyberScan pH5500/ 5000 It is a well known fact that pH is a temperature dependent measurement. The factory default setting is 25. if you are taking the pH of a solution that is not 25 and you are not using an Automatic Temperature Compensation (ATC)( probe, then you should meter the temperature value of that solution in order to get the correct pH value.
Instruction Manual CyberScan pH5500/ 5000 6.9 Set Isopotential Point Set pH Isopotential Point enter 0.0 mV Current ISO POINT clear Touch clear to delete current Iso Point numeric touchpad to input the new Iso Point and then touch enter to accept...
Page 61
Instruction Manual CyberScan pH5500/ 5000 The Isopotential point is the millivolts reading for an electrode at which temperature has no effect on the measurement. pH electrodes are constructed so that the isopotential point is theoretically zero millivolts. This is very close to a pH of 7. Most pH electrodes do not achieve this value precisely.
Instruction Manual CyberScan pH5500/ 5000 6.10 Set Alarm Limits Set pH Limits Current LIMITS pH Alarm pH Minimum 0.00 pH Maximum 14.00 edit Set mV Limits Set Ion Limits Current LIMITS Current LIMITS Ion Alarm mV Alarm mV Minimum -1800.0 Ion Minimum 0.00 Ion Maximum...
Page 63
Instruction Manual CyberScan pH5500/ 5000 This option allows you to set alarm limits for the pH measuring mode. If the pH value of the measurement is outside the boundaries set by the minimum and maximum limits, an audible alarm and/ or a visual warning will appear to let you know that your sample measurement was outside of the set limits.
Instruction Manual CyberScan pH5500/ 5000 6.11 Set Print Criteria Set pH Print Criteria Set mV Print Criteria Current PRINT CRITERIA Current PRINT CRITERIA - Date/Time/Channel - Date/Time/Channel - Sample ID# - Sample ID# - pH Measurement - mV Measurement - Temperature - ATC - Temperature - ATC - Last Standardization -- Meter Model # / serial #...
Page 65
Instruction Manual CyberScan pH5500/ 5000 This screen allows you to select which criteria are printed with the measurement when you print the data or send it to the computer. The status of the current print criteria is displayed on the screen. The criteria option is active if “ON”...
Instruction Manual CyberScan pH5500/ 5000 6.12 Set Print Interval Set pH Print Interval enter manual Current INTERVAL timed You have three options for setting the print interval: manual printing, stable reading printing, and timed interval printing. For manual printing of data In this mode, data is printed only when you touch print on the pH (mV, Ion) Measure screen.
Page 67
Instruction Manual CyberScan pH5500/ 5000 For stable reading printed In this mode, data is printed every time the meter recognizes the current pH (mV, Ion) measurement as stable. 1. Access the Set Print Interval screen from the pH (mV, Ion) Setup screen.
Instruction Manual CyberScan pH5500/ 5000 6.13 Set Data Storage Criteria Set pH Data Storage Criteria Set mV Data Storage Criteria Current DATA STORAGE CRITERIA Current DATA STORAGE CRITERIA - Date/Time/ Channel - Date/Time/ Channel - Sample ID# - Sample ID# - pH measurement - mV measurement - Temperature - ATC...
Page 69
Instruction Manual CyberScan pH5500/ 5000 This screen allows you to select what criteria are stored in the meter’s memory with the measurement when you save the data. Data is stored only if a Sample ID# has been assigned. The status of the current data storage criteria is displayed on the screen.
Instruction Manual CyberScan pH5500/ 5000 6.14 Set Display Resolution Set pH Display Resolution enter Current RESOLUTION X.XX X.XX Select desired display resolution and then touch enter to accept X.XXX...
Page 71
Instruction Manual CyberScan pH5500/ 5000 This mode allows you to set the display resolution that you desire on the screen. You have the choice of one, two or three decimal places. To Set Display Resolution 1. Access the set display resolution screen from the pH Setup screen.
Instruction Manual CyberScan pH5500/ 5000 6.15 Set Display Configuration Set pH Display Configuration Set mV Display Configuration Current DISPLAY CONFIGURATION Current DISPLAY CONFIGURATION - Last Standardization - Date - Date - Time - Time - measurement channel - measurement channel - sample ID# - sample ID# - temperature...
Page 73
Instruction Manual CyberScan pH5500/ 5000 This function will allow you to choose what information you would like to be displayed on the pH Measure screen, particularly the information contained in the data box at the bottom of that screen. To Set Display Configuration 1.
Instruction Manual CyberScan pH5500/ 5000 6.16 View Stored Data View Stored Data Data POINTS Sample ID Date Operator enter arrow keys to highlight desired sort option and then touch enter to accept...
Page 75
Instruction Manual CyberScan pH5500/ 5000 This meter has memory capacity of up to 250 data points. The View Stored Data screen allows you to sort and look at specific data points. The stored data can be sorted by sample identification number, date or operator identification number.
Page 76
Instruction Manual CyberScan pH5500/ 5000 To sort by Date 1. Access the Date sort option from the View Stored Data screen. 2. Touch clear to delete the current date. 3. Use the numeric keypad to enter the date on which the data points you want to view were saved.
Page 77
Instruction Manual CyberScan pH5500/ 5000 If an operator ID is entered and no data points are stored with that operator ID, you will see a message indicating the operator ID was not found. Touch OK to return to the operator ID keypad and enter a new operator ID.
Instruction Manual CyberScan pH5500/ 5000 7.1 Access mV Setup E U T E C H Setup INSTRUMENTS CyberScan pH 5500 Select from the options to the right Select from the options to the right setup system Last std: Jan 15 @ 11:11 am There are two ways to access the mV mV Setup (channel 1) Setup screen.
Page 80
Instruction Manual CyberScan pH5500/ 5000 Measure E U T E C H INSTRUMENTS abs mV channel 1 meas CyberScan pH 5500 Touch meas to measure sample Select from the options to the right Touch for Relative mV reading setup January 15, 2002 11:11am 00000 25.0°C...
Page 81
Instruction Manual CyberScan pH5500/ 5000 To Set Sample ID# When this option is active, each time you touch print on the Measure screen the mV value along with the date/time/channel and the sample ID# will be sent to data storage. You can manually enter an alphanumeric identification number of up to 10 characters for any sample, or you can have the meter sequentially number your samples beginning at the number of your choice.
Page 82
Instruction Manual CyberScan pH5500/ 5000 To Set Print Interval You have three options for setting the print interval: manual printing, stable reading printing and timed interval printing. The mV Print Interval is set the same way as previously described in pH setup.
Instruction Manual CyberScan pH5500/ 5000 7.2 Set Display Resolution Set mV Display Resolution enter Current RESOLUTION X.XX X.XX Select desired display resolution and then touch enter to accept...
Page 84
Instruction Manual CyberScan pH5500/ 5000 To Set Display Resolution This mode allows you to set the display resolution that you desire on the screen. You have the choice of zero or one decimal place. 1. Access the set display resolution screen from the mV setup screen.
Instruction Manual CyberScan pH5500/ 5000 ION SETUP Ion Setup (channel 1) ISE SETUP OPTIONS - Set Sample ID# - Select Ion Method - Select Measurement Units - Select Auto Read Mode - Set Ion Stability Criteria - Set Default Temperature edit - Set Electrode Type...
Instruction Manual CyberScan pH5500/ 5000 8.1 Access Ion Setup E U T E C H Setup INSTRUMENTS CyberScan pH 5500 Select from the options to the right Cond Cond Select from the options to the right setup system Last std: Jan 15 @ 11:11 am Ion Setup (channel 1) There are two ways to access the Ion Ion SETUP OPTIONS...
Page 87
Instruction Manual CyberScan pH5500/ 5000 Measure E U T E C H INSTRUMENTS channel 1 meas CyberScan pH 5500 STANDARDS Last std: Jan 15 @ 11:11 am setup Touch to access standardize mode Select from the options to the right Cond setup January 15, 2002...
Instruction Manual CyberScan pH5500/ 5000 8.2 Select Ion Method To Set Sample ID# When this option is active, each time you touch print on the Measure screen the ion concentration value along with date/ time/ channel and the sample ID# will be sent to data storage. You can manually enter an alphanumeric identification number of up to 10 characters for any sample or you can have the meter sequentially number your samples beginning at the number of your choice.
Page 89
Instruction Manual CyberScan pH5500/ 5000 Direct Reading with Standards The direct reading method is the most common method of ion measurement. When using this method, you will use standards of known values to standardize the meter with ion standards. An ionic strength adjuster (ISA) is added to both the standards and the sample.
Instruction Manual CyberScan pH5500/ 5000 Analate Subtraction The analate subtraction method is particularly useful in the event that there is no ion specific electrode available to directly measure the concentration of the ion of interest. This method employs the addition of a sample of the ion of interest to a standard containing a different ion of known concentration.
Page 91
Instruction Manual CyberScan pH5500/ 5000 To Select Auto Read Mode You can use this meter when the Auto Read Mode function is active or when it is inactive. When the Auto Read function is active, the meter will lock onto a reading when the meter recognizes it as stable. The meter will not deviate from this reading until meas is touched.
Instruction Manual CyberScan pH5500/ 5000 8.4 Select Electrode Type This screen allows you to select the type of electrode that is specific for your ion of interest. Your choice of electrode informs the meter which default slope to use in the calculations made for the electrode efficiency and sample concentration.
Page 93
Instruction Manual CyberScan pH5500/ 5000 To Set Isopotential point The Isopotential point is the millivolts reading for an electrode at which temperature has no effect on the measurement. Ion selective electrodes, unlike pH electrodes, are not designed to achieve any particular isopotential point.
Instruction Manual CyberScan pH5500/ 5000 8.5 Set Significant Digits Set Ion Significant Digits enter Current SIG DIGITS Select desired significant digits and then touch enter to accept Example of 1 significant digit Example of 2 significant digits channel 1 channel 1 Example of 3 significant digits channel 1...
Page 95
Instruction Manual CyberScan pH5500/ 5000 This option allows you to select the number of significant figures that will be used when reporting your ion measurement. You have the choice of one, two or three significant digits. To Set Significant Digits 1.
Instruction Manual CyberScan pH5500/ 5000 pH OPERATION Measure channel 1 7.00 meas Buffers Last std: Jan 15 @ 11:11 am meas to measure sample Touch Touch to standardize January 15, 2002 11:11am 00000 25.0°C auto buffer slope 100% 0000.0...
Page 97
Instruction Manual CyberScan pH5500/ 5000 In this mode, you will able to measure the pH of a wide variety of samples. Before measuring pH, you will need to standardize the meter using buffers with known pH values. It is good practice to standardize the meter frequently using a minimum of two buffers.
Instruction Manual CyberScan pH5500/ 5000 9.1 Standardizing Measure channel 1 7.00 meas STABLE Buffers Last std: Jan 15 @ 11:11 meas to measure sample Touch Touch to standardize January 15, 2002 11:11am 00000 25.0°C auto buffer slope 100% 0000.0 Standardize channel 1 7.00 meas...
Page 99
Instruction Manual CyberScan pH5500/ 5000 To Standardize the meter with Auto Buffer Recognition 1. Touch STD on the pH measure screen to access the standardized screen 2. Touch clear to delete a previous standardization. All pH standardization values will be cleared. The meter will remain on the standardization screen at this point.
Page 100
Instruction Manual CyberScan pH5500/ 5000 Standardize channel 1 7.00 meas Buffers Last std: Jan 15 @ 11:11 Current BUFFER GROUP clear to delete previous buffers Touch Insert electrode into buffer and stir Touch to standardize new buffer January 15, 2002 11:11am 00000 25.0°C...
Page 101
Instruction Manual CyberScan pH5500/ 5000 To standardize the meter with Manual Buffer Recognition 1. Touch std on the pH measure screen to access the standardize screen 2. Touch clear to delete a previous standardization. All pH standardization values will be cleared. The meter will remain on the standardization screen at this point.
Instruction Manual CyberScan pH5500/ 5000 9.2 pH Measurements Measure channel 1 7.00 meas Last std: Jan 15 @ 11:11 am Buffers meas to measure sample Touch Touch to standardize January 15, 2002 11:11am 00000 25.0°C auto buffer slope 100% 0000.0...
Page 103
Instruction Manual CyberScan pH5500/ 5000 The measure screen provides readout of the current sample measurement. You can use this meter when the auto read function is active or when it is inactive. When the auto read function is active, the meter will lock onto a reading when the meter recognizes it as stable.
Instruction Manual CyberScan pH5500/ 5000 10 mV OPERATION This mode is used to measure oxidation/ reduction potential (ORP/redox), perform titration and to verify the function of the meter. The mV measure function allows you to continuously monitor the mV potential of the electrodes in use. This can be done in either absolute or relative mV.
Instruction Manual CyberScan pH5500/ 5000 10.1 mV Measurement In the mV mode, you will be able to make measurements in either absolute or relative millivolts, access the mV Setup screens and print your results to a printer or a computer. Connect the electrodes you will be using to the meter.
Instruction Manual CyberScan pH5500/ 5000 11 ION OPERATION Measure - Ion channel 1 meas 10.0 STNDS Last std: Jan 15 @ 11:11 am meas to measure sample Touch Touch to access standardize mode or measure blank January 15, 2002 11:11am 00000 25.0°C slope...
Page 107
Instruction Manual CyberScan pH5500/ 5000 This CyberScan meter offers you unparalled versatility with respect to ion analysis. You can choose one of the 6 methods of analysis, the specific electrode you will be working with and also the units that you would like the meter to use when reporting the ion concentration.
Instruction Manual CyberScan pH5500/ 5000 11.1 Direct Reading Methods Standardize - Ion Standardize - Ion channel 1 channel 1 - - - - - - meas meas 10.0 setup setup Last std: Jan 15 @ 11:11 Not standardized STND STND h clear to delete previous standards Touc clear...
Page 109
Instruction Manual CyberScan pH5500/ 5000 There are two direct reading methods. Direct Reading with standards and Direct Reading with Blank Offset. These types of measurements allow you to directly read the concentration of your sample after standardizing the meter with ion standards of known values or blank values.
Page 110
Instruction Manual CyberScan pH5500/ 5000 Measuring ion concentration in the Direct Reading with Standards Mode With Auto Read on 1. Add Ionic strength adjuster to the sample. 2. Immerse the ion selective electrode in the stirring sample. 3. Touch meas to begin measuring your sample. The meter will accept the reading and display STABLE when the measurement meets the selected stability criteria.
Page 111
Instruction Manual CyberScan pH5500/ 5000 Standardize - Ion Standardize - Ion channel 1 - - - ISE Standardize meas meas You have selected Direct Reading with Blank offset operation. STND Is this reading a STANDARD or BLANK? setup setup STND Not Standardized clear clear...
Page 112
Instruction Manual CyberScan pH5500/ 5000 Direct Reading with Blank Offset This method presents you with the ability to measure samples with low level concentrations of the ion of interest. It eliminates background interference by subtracting the value of a blank concentration from the value for the sample.
Page 113
Instruction Manual CyberScan pH5500/ 5000 If you touched STND 8. Use the keypad to enter the value of the standard. 9. Touch enter to accept the value. The meter will accept the standard value when it recognizes the reading as stable. 10.
Instruction Manual CyberScan pH5500/ 5000 11.2 Incremental Methods Standardize - Ion Standardize - Ion channel 1 channel 1 - - - - - - meas meas 10.0 setup setup STND Last std: Jan 15 @ 11:11 STND Not standardized Touch clear to delete previous standards clear clear Insert electrode into standard and stir...
Page 115
Instruction Manual CyberScan pH5500/ 5000 The incremental methods which include known addition, known subtraction, analate addition and analate subtraction and are particularly useful with samples containing complex ionic backgrounds which cannot be matched in standards. They are also useful with occasional samples whose temperatures vary. More information on these methods is available in the ISE Theory Section (See page 133).
Instruction Manual CyberScan pH5500/ 5000 11.3 Known Addition Method Measure - Ion Measure - Ion channel 1 channel 1 1.00 10.0 setup setup Not standardized STND STND Last std: Jan 15 @ 11:11am Insert electrode into standard and stir Touch to access standardize mode Touch to standardize new standard...
Page 117
Instruction Manual CyberScan pH5500/ 5000 Standardize - Ion Standardize - Ion channel 1 channel 1 ISA Volume Concentration meas meas STABLE STABLE STND STND Not Standardized Not Standardized setup setup clear clear Touch meas to measure sample Touch meas to measure sample Touch to access standardize mode Touch...
Page 118
Instruction Manual CyberScan pH5500/ 5000 Measuring Ion Concentration in Samples Using the Incremental Methods When performing known addition, you will add an aliquot of standard solution containing the ionic species of interest to a known volume of your sample. The mV readings of the sample can then be calculated based upon the difference in the mV readings.
Page 119
Instruction Manual CyberScan pH5500/ 5000 To take a measurement 1. Immerse the ion selective electrode in a known volume of the sample solution. 2. Stir the sample. 3. Touch KA to take the first mV reading of sample. Wait for the reading to stabilize (STABLE will appear).
Instruction Manual CyberScan pH5500/ 5000 11.4 Known Subtraction Method The procedure for known subtraction is very similar to that for known addition method. An aliquot of standard is added to a known volume of the sample. The difference is that the standard does not contain the same ionic species that you are trying to measure in the sample.
Page 121
Instruction Manual CyberScan pH5500/ 5000 10. Use the keypad to enter the concentration of the standard solution added to the sample. 11. Touch enter to accept the value. 12. Use the keypad to enter the concentration of the standard solution added to the sample 13.
Instruction Manual CyberScan pH5500/ 5000 11.5 Analate addition The analate addition method is useful when your sample is at an elevated temperature and you want to negate the effect that temperature will have on the concentration calculation. This method is also useful if you have only a small amount of sample available. The analate addition method differs only slightly from the known addition method.
Page 123
Instruction Manual CyberScan pH5500/ 5000 10. Use the keypad to enter the concentration of the standard solution added to the sample. 11. Touch enter to accept the value. 12. Use the keypad to enter the concentration of the standard solution added to the sample 13.
Instruction Manual CyberScan pH5500/ 5000 11.6 Analate Subtraction The analate subtraction method is useful when no ion specific electrode is available to measure the ion of interest directly. Like known subtraction the sample and the standard contain different ionic species. However, the ion being measured is in the standard and not in the sample.
Page 125
Instruction Manual CyberScan pH5500/ 5000 10. Use the keypad to enter the concentration of the standard solution added to the sample. 11. Touch enter to accept the value. 12. Use the keypad to enter the concentration of the standard solution added to the sample 13.
Instruction Manual CyberScan pH5500/ 5000 CLEANING The touch screen should be kept as clean as possible to preserve optical properties. Attempt to keep the screen free of dirt, dust fingerprints, etc. long term contact with abrasive materials will scratch the surface, and impair image quality. To clean, use a damp nonabrasive cloth towel and any commercially available window cleaner.
Instruction Manual CyberScan pH5500/ 5000 13 TROUBLE SHOOTING Your CyberScan meter contains many error messages to provide aid should trouble occur with a measurement or meter operation (touch pad and input errors). The messages are accompanied by a brief description of the error, and in some cases advice on how to correct it.
Eutech Instruments-supplied power supply must be used. Use of a power supply that is not approved by Eutech Instruments may cause safety hazards and/ or cause unit to exceed EMC limits and/ or damage unit. When using this...
Instruction Manual CyberScan pH5500/ 5000 15 APPENDIX 15.1 Determining Isopotential Points Experimentally The isopotential point of an electrode system is the point at which electrode potential is unaffected by a change in temperature. The coordinates of this point would be coincident with the system’s Zero Potential Point (pX iso, In practical systems, however, this coincidence rarely occurs, and for...
Instruction Manual CyberScan pH5500/ 5000 15.2 Data Management Your CyberScan meter is equipped with an RS-232 output port for sending measurement readings to a printer or computer. To use the meter with a printer or a computer, the printer or computer must have an RS-232 part, also.
Instruction Manual CyberScan pH5500/ 5000 15.3 Factory Default Settings The following is the list of factory default settings for this CyberScan meter. You can reset your meter to the factory default settings by accessing the Reset to Factory Defaults from the System Setup screen.
Page 132
Instruction Manual CyberScan pH5500/ 5000 Mode Screen Default Setting Display configuration All parameters ON Print configuration All parameters ON Data configuration All parameters ON Sample ID# None Isopotential Point 0 mV Print interval manual Display resolution 2 significant digits Auto Read Mode Temperature Units Celsius Default Temperature...
Instruction Manual CyberScan pH5500/ 5000 15.4 pH Theory Since its introduction by the Danish Chemist Sorensen in 1909, pH measurement has become one of the most commonly used and important measurements in both laboratory and industry settings. pH measurement and control is vital to a wide array of endeavors including municipal and industrial wastewater treatment, and textile, pharmaceutical, food and pH.
Page 134
Instruction Manual CyberScan pH5500/ 5000 At 25 °C in pure water, the concentration of hydronium ions is extremely small, 1 X 10 Moles/liter, and balanced by an equal concentration of hydroxyl ions.. The equilibrium constant, Kw of water is the product of the hydronium ion and hydroxyl ion concentrations: Kw= [H ] [OH ] = [1 X 10...
Page 135
Instruction Manual CyberScan pH5500/ 5000 pH Measurement The pH value of a sample can be determined in several ways. These include the use of organic dyes which change colour in certain pH ranges. The dyes can be added directly to the solution or impregnated onto paper which may be dipped into the solution.
Page 136
For some pH glasses, this means an error as high as 0.5 units in highly alkaline, high-sodium solutions. The special FS-5 glass using in Eutech Instruments electrodes exhibits must less sodium error in these solutions. < 0.05 pH units.
Page 137
Instruction Manual CyberScan pH5500/ 5000 Reference electrodes typically consist of three elements: an internal reference electrode, a filling solution, and a reference junction through which the filling solution can “flow”, and provide electrical contact with the sample and the internal reference electrode. The most convenient internal reference electrodes are made from a metal (such as silver or mercury) and its sparingly soluble salt (silver chloride or mercurous chloride).
Instruction Manual CyberScan pH5500/ 5000 15.5 ISE Theory Ion selective electrodes (ISE) respond to ionic activity or concentration in a liquid because of the characteristics of the electrode’s sensing membrane. Ideally, the ISE develops an electrical potential which is proportional to the activity or concentration of the ion which the membrane is selective.
Page 139
Instruction Manual CyberScan pH5500/ 5000 This form of the Nernst equation states that the electrode potential varies directly with the log of the concentration in straight line manner, and results are then expressed in the preferred concentration units. The slope of the line is equal to the value of: Slope = S = 2.3RT/ nF The table below gives theoretical slope values at 25°C Slope...
Page 140
Instruction Manual CyberScan pH5500/ 5000 Known Addition and Known Subtraction With the Known Addition and Known Subtraction methods, the initial millivolts output of the electrode in the sample solution is measured. Then an aliquot of standard solution is added, and a final millivolts potential is measured.
Page 141
Instruction Manual CyberScan pH5500/ 5000 Analate Addition and Analate Subtraction In the Analate methods, the initial millivolt reading is taken with the electrode immersed in the standard; then an aliquot of sample is added, and the final millivolt reading is taken. Sample concentration is calculated from the change in the millivolt readings.
Instruction Manual CyberScan pH5500/ 5000 16 METER SPECIFICATIONS Feature Details pH 5500 pH5000 640X480 digit LCD Display Screen size 4½” X 6” Measurement display ¾“ height Temp/etc. display height ¼” Menu options Extensive Help screens Extensive Configurable display Keypad controls Context specific touch screen Memory...
Page 143
Instruction Manual CyberScan pH5500/ 5000 Feature Details pH 5500 pH5000 Temperature Mode Range -5.0 to +105.0°C Resolution 0.1°C Accuracy ±0.2°C General Inputs/ outputs BNC, 2Pin, 2ATC, RS232, DIN (for FET) Electrical requirements Power Adapter 115VAC/ 12VDC, (Depends (Depends 230 VAC/ 12VDC on Order on Order Code)
Instruction Manual CyberScan pH5500/ 5000 17 ACCESSORIES Consult your Authorized Distributors for these items and other range of specialized pH electrodes or Ion Selective Electrodes. 17.1 Replacement Meters and Adapters CODE NO. DESCRIPTION EC-PH5500/11S CyberScan pH 5500 Bench pH/ Ion/ mV/ RS232 (dual channel) with integral electrode stand and 110 VAC power adapter, 2- flat pin US type (center negative) EC-PH5500/21S...
Page 145
Instruction Manual CyberScan pH5500/ 5000 17.2 General Purpose pH / ORP Electrodes TYPICAL CODE NO. DESCRIPTION APPLICATIONS EC-FC725522-01B Double junction pH combination, General purpose, sealed polymer gel, Ag/AgCl rugged use reference EC-FC725522R-01B Double junction pH combination, General purpose, refillable electrolyte, Ag/AgCl rugged use reference EC-FG73504-01B...
Page 146
Instruction Manual CyberScan pH5500/ 5000 17.3 Specialty pH Electrode TYPICAL CODE NO. DESCRIPTION APPLICATIONS EC-620-297 MicroProbe Glass body Small samples, Tris EC-620-116 Flow & Flushable reference junction, General purpose, plastic body combination pH Tris, soils, sludges, electrode viscous EC-620-109 Flow & Flushable reference DOUBLE General purpose, junction, plastic body combination pH Tris, soils, sludges,...
Page 148
Instruction Manual CyberScan pH5500/ 5000 WARRANTY Eutech Instruments supplies the bench meters with a 3-year warranty, probes with 6-month warranty, against manufacturing defects from the date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warranty period, please return, freight prepaid, and correction will be made without charge.