Connecting Peripherals - Thermo Scientific ALPHA PH 550 Manual

Monitor ph / orp
Table of Contents

Advertisement

2.3

Connecting Peripherals

2.3.1
Connection Terminals
Remove Back Cover:
Remove the screws from the four corners at the back of the pH Monitor. Remove
the back cover. The connectors are exposed on the back PCBA as shown in the
Figure 1 below.
Connectors:
J11 – 24V DC power
J8 – 9V DC power
J10 - pH electrode & Temperature probe connections (wiring has to be done in
J11 Connections
1.
+ve
2.
Gnd
J10 Connections
1.
pH Sense
2.
pH Reference
3.
No Connection
4.
No Connection
5.
Pt 100 Compensate
6.
Pt 100 Sense
7.
Pt 100 GND
8.
Potential Matching Pin
the detachable connector
1)+v
e
Figure 1: Outer Side of Back PCBA
24V DC
J1
9V DC Power
Screw
J10
- 5 -