2.3
Connecting Peripherals
2.3.1
Connection Terminals
Remove Back Cover:
Remove the screws from the four corners at the back of the pH Monitor. Remove
the back cover. The connectors are exposed on the back PCBA as shown in the
Figure 1 below.
Connectors:
•
J11 – 24V DC power
•
J8 – 9V DC power
J10 - pH electrode & Temperature probe connections (wiring has to be done in
the detachable connector
J11 Connections
1.
+ve
2.
Gnd
J10 Connections
1.
pH Sense
2.
pH Reference
3.
No Connection
4.
No Connection
5.
Pt 100 Compensate
6.
Pt 100 Sense
7.
Pt 100 GND
8.
Potential Matching Pin
24V DC
1)+v
J1
e
J10
Figure 1: Outer Side of Back PCBA
- 5 -
J8
9V DC Power
Screw